Sana comfort 150 and 250 REHA
Article no.: 2.520064 rev.: b
Page 8 of 24
3
Installation
3.1
Location
Install the device in a suitable position (refer to safety instructions in section 5).
The unit should not be stored or operated in wet, moist or dusty surroundings. Nor should the unit be
exposed to direct sunlight or other sources of warmth.
The unit should not come into contact with acidic vapours or fluids.
The unit should not be placed near X-ray units, large transformers or electrical motors. There must be a
distance of at least one meter between the unit and the mains network.
3.2
Assembly instructions
3.2.1
Unpacking and assembling
Install the control panel after unpacking the unit. In order to do so, insert the two tabs on the back of the
control panel into the handle bar pipe and press them downward to their stop. The operator's side should
usually face the front so that the display can be seen by the person operating the machine. Connect the
potential equalisation cable to the flat plug at the back of the control panel.
Connect the main plug with the connecting socket. Fasten the rear cover with 4 screws.
Remove the seat's transport protection.
With the help of the base adjustor on the rear lower side of the ergometer, adjust the device so that it is
flush with the floor. The ergometer is then fully stabilised.
3.2.2
Seat adjustment
Mechanical seat adjustment
The seat is infinitely variable horizontally in order to accurately adjust the distance to the pedals. The
seat can be adjusted for persons between 150 cm and 210 cm in height. The handle next to the control
console ensures safe grip while positioning the feet on the pedals.
A ball-grip lever is used to adjust the sitting position. This lever is located on the right hand side, under-
neath the seat, and is easily accessible. Press the ball handle downwards to unlock the seat break, and
adjust the seat until the correct sitting position has been found. Pull the lever upwards as far as it will go
to lock the seat.
The seat construction is designed for patients weighing up to 250 kg.
Motorised seat adjustment
Adjust the seat by use of the arrow buttons.
3.2.3
Connecting the blood pressure cuff
The interfaces for the air tube and the microphone are located underneath the seat's guiding rail. Con-
nect the blood pressure cuff by screwing the hose connection onto the joining nipple and inserting the
microphone plug in the socket next to it.
3.2.4
Mains connection
Establish potential equalisation (see section 2.3) and plug the supplied power cable into an earthed
socket. As the device is preset to the local mains voltage (refer to section 8.4), you can switch it on using
the main switch at the front.