9 Technical Specifications
48 |
48
201000401000 / Rev 04
9.2 Application Software ers2
9.3 Receiver
9.3.1
Summary
Description
Data
Heart rate calculation
The heart rate is calculated by the ers2 software which
uses the
Pan
Tompkins
algorithm for QRS detection.
Tall T-wave rejection capability
For T-waves <= 0.8 mV, the heart rate indication is
within the specified error limits (+/– 10 % or 5 BPM)
Response to irregular rhythm
A1 / (3a):
40 BPM
A2 / (3b):
30 BPM
A3 / (3c):
120 BPM
A4 / (3d):
90 BPM
to IEC 60601-2-27
Response time to changes in heart
rate
80 BPM increase to 120 BPM: max. 6 s
80 BPM decrease to 40 BPM:
max. 12 s
Time to alarm for tachycardia
B1: 11 s
0.5
B1: 12 s
2
B1: 23 s
B2: 11 s
0.5
B2: 12 s
2
B2: 16 s
To IEC 60601-2-27
Visual alarm
Location: ers2 software monitoring screen
Color/Modulation: 1-Hz modulation between red and
yellow
Size: 160
23 pixels
[Monitor resolution 1920
1080]
Visual alarm update: every 2 min
Audible alarm
Source: PC sound card
Frequency of sound: 1 Hz [on/off]
Audible alarm update: every 2 min
Description
Data
Transmission mode
Bluetooth 2.1 + EDR, class 1
Frequency range
2402–2483.5 MHz, ISM band
Operating mode
continuous operation
Power supply
via USB current (max.): 175 mA
Protection class
SK II
Degree of protection
IP 20
Max. output power
18.0 dBm