•   Meter Dimension    : 156x73x35mm(Lx W x H).

 

•   Probe  Dimension  : 

                           Wire length   : 160cm 

                           Probe length : 128cm 

                           Probe diameter of tip : 1.6cm   

                           Probe diameter of base : 3.2cm 

•   Meter weight        : 250g.(without batteries) 

•   Probe  Weight     : 240 g. 

•   Power supply       : UM-4(AAA)1.5V*6PCS or 

                               AC100~240V DC9V/0.5A(9mm). 

•   Battery  life           : 10 hours. 

•   Operating T

  emperature & Humidity: 

                            -20°C to +50°C, <80 %RH. 

•   Storage Tempe

  rature & Humidity :   

                          -10°C to +50°C, <70 %RH. 

•   Data outpu

  t         :  USB PC serial interface. (TM-4003) 

•   Display                 :  60mm(W) x 47mm(L) monochrome 

                                LCD and backlight. 

•   Standard Accessories : 

   UM-4  (AAA)1.5V battery  X  6  Pcs

  

       

 

 

33.0%RH. 

The duration of calibration time is 10 minutes. 

After end of the 33.0%RH calibration, you can 

press 

  button to shut down or press 

 

button to into 75%RH calibration. 

 

D. Humidity calibration (75%RH) 

Put the probe in the 75%RH chamber and 

press 

  button to select the 75%RH and 

press 

  button to start 75%RH calibration. 

The duration of calibration time is 10 minutes. 

After end of the 75%RH calibration, then will be 

into the normal display. 

 

E. Temperature calibration (25.0 ) 

1. Press 

  button to select the 25.0  and 

press 

  button to start 25.0

 calibration. 

2. Confirm with the standard value of the error, 

press the 

  and 

  button to adjust the 

temperature and the display will flash. 

3. Press 

  to complete the adjustment 

and into the normal display. 

 

19.  PRECAUTIONS 

•   Keep the meter away from electromagnetic 

Содержание EQ-TM- 4001

Страница 1: ...correct specification of the battery and correctly installed in the battery holder 25 END OF LIFE Caution This symbol indicates that equipment and its accessories shall be subject to a separate colle...

Страница 2: ...obe Instant AVG 2 3V MAX flow measurement Max A VG Min data hold function Low batter indication LCD triple display with LED backlight Manual data memory and read function 99 recor ds TM 4001 Auto pow...

Страница 3: ...rned when you have finished testing the high wind speed more than 20m s and remove the probe To extend the probe hold the handle in one hand while pulling on the probe tip with the other hand Do not h...

Страница 4: ...bration time is 10 minutes After end of the 33 0 RH calibration you can press button to shut do wn or press button to into 75 RH calibration D Humidity calibration 75 RH Put the probe in the 75 RH cha...

Страница 5: ...s button to select the 10 00m s and press button to st art calibration 10m s The duration of calibration time is 5 minutes After end of the 10m s calibration then will be into the normal display C Hum...

Страница 6: ...more than 2 seconds to disable this feature 18 CALIBRATION Please shut down the meter first then hold and press button and pr ess button to enter into calibration mode It will show 0 00 m s on the up...

Страница 7: ...le or disable hold function The HOLD function will be unavailable if calibration function is enabled RH humd i ity indication C Degree Celsius C Degree Fahrenheit hPa Hectopascal indication inHg Inch...

Страница 8: ...ee the n SL display on the LCD that means the auto power off function is off If you want enable auto power off please restart the meter 8 BACKLIGHT Press buttonto turn on the backlight pres this butto...

Отзывы: