Troubleshooting
Troubleshooting
29
Figure 3-3. The lamp does not light up.
Figure 3-4. Scanned image is unclear.
Start
Is connector
CN6 on the main
circuit board
connected?
Are connectors
CN1 and CN2 on
the CCD circuit
board connected?
Is fluorescent lamp
inserted securely
into inverter circuit
board?
Is the harness
connected to the
main circuit board
and carriage?
Replace the main
circuit board.
Replace the
carriage.
Reset into inverter
circuit board.
Connect the
connector.
Connect the
connector.
Start
Is the document
platen dirty?
Is the mirror inside
the carriage dirty?
Is the
Sheet White section
dirty or bent?
Is the
CCD sensor circuit
board normal?
Is the scanned
image okay?
Mechanism problem
(replacement required)
Replace the main
circuit board.
Replace the
carriage unit.
Replace the upper
housing.
Clean.
(See 6.1.1 Cleaning.)
* White spots appear in image.
* Vertical stripes appear in image.
Clean.
(See 6.1.1 Cleaning.)
Содержание Perfection 3200 Photo
Страница 1: ...EPSON Perfection3200Photo Color Image Scanner SESC02 003 SERVICE MANUAL ...
Страница 5: ...Revision Status Revision Issued Date Description A November 18 2002 First Release ...
Страница 8: ...C H A P T E R 1 PRODUCT DESCRIPTION ...
Страница 21: ...C H A P T E R 2 OPERATINGPRINCIPLES ...
Страница 29: ...C H A P T E R 3 TROUBLESHOOTING ...
Страница 36: ...C H A P T E R 4 DISASSEMBLYANDASSEMBLY ...
Страница 58: ...C H A P T E R 5 ADJUSTMENT ...
Страница 61: ...C H A P T E R 6 MAINTENANCE ...
Страница 64: ...C H A P T E R 7 APPENDIX ...
Страница 70: ......
Страница 71: ......
Страница 72: ......