EnviteC SpO2 sensors and accessories
The SpO2 sensors are transmission sensors and contain two LEDs as well as a photodiode for this spectrum.
Wavelengths
•
Reusable sensors: 660/890 nm at 3.5 - 4.5 mW
•
Disposable sensors: 660/890 nm at 2.5 - 3.5mW
The sensors are individually detected by MySign
®
S for the best possible measurement precision.
7.1 Selection and application of a sensor
Before using the sensors, carefully read the associated operating manual as well as all warnings
and other instructions.
Never use damaged sensors. Never use a sensor with unprotected optical components.
Only use sensors approved by EnviteC for SpO2 measurements.
When selecting the sensor, consider the weight and activity level of the patient. Also evaluate
whether there is sufficient blood circulation at the application site.
The sensor should be protected from heavy exposure to light since this can cause measurement
errors. If the signal quality is not sufficient after about 10 seconds, try affixing the sensor at a
different application site.