Pulse oximeter MySign
®
S
The artifact leveling (AL) is used to suppress movement artifacts for the SpO2 and pulse rate
parameters. In addition, the pulse rate is checked for plausibility with a deviation suppression (DS)
function.
The MySign
®
S is calibrated to pulse-oximetric hemoglobin oxygen saturation with dyshemoglobin-
free blood based on reference measurements via fractional saturation measurement (CO-oximeter).
If the level of dysfunctional hemoglobin is high, the precision of the measurement is impaired.
Intravascular dyes, nail polish and artificial fingernails can also impair the precision of the
measurement.
In addition, the measurement precision can be impaired by strong ambient light or direct sunlight. In
such cases, the sensor can be covered.
If you doubt the precision of the measurement, check the vital signs of the patient with the help of
other methods. Then make certain that the MySign
®
S is functioning properly.