48
Burst Frequency
: 1 – 9 Hz, in steps of 1 Hz
Amplitude
: 0 - 140 mA
Symmetrical
Phase Duration
: 10 - 400 µs in steps of 5 µs
Pulse Frequency
: 1 - 200 Hz, in steps of 1 Hz
Frequency Modulation (spectrum)
: 0 - 180 Hz in steps of 1 Hz
Modulation Program
: 1/1, 6/6, 12/12, 1/30/1/30 s
Surge Program
: Yes
Amplitude
: 0 - 140 mA
Burst Symmetrical
Phase Duration
: 10 - 400 µs in steps of 5 µs
Pulse Frequency
: 1 - 200 Hz, in steps of 1 Hz
Burst Frequency
: 1 – 9 Hz, in steps of 1 Hz
Amplitude
: 0 - 140 mA
Premodulated
Carrier Frequency
: 2, 2.5, 3, 3.5, 4, 5, 6, 7, 8, 9, 10 kHz
Beat Frequency
: 0 - 200 Hz in steps of 1 Hz
Frequency Modulation (spectrum)
: 0 - 180 Hz in steps of 1 Hz
Modulation program
: 1/1, 6/6, 12/12, 1/30/1/30 s
Surge Program
: Yes
Amplitude
: 0 - 140 mA
Russian Stimulation
Carrier Frequency
: 2, 2.5, 3, 3.5, 4, 5, 6, 7, 8, 9, 10 kHz
Burst frequency
: 0 - 100 Hz in steps of 1 Hz
Burst / Pause
: 1:1, 1:2, 1:3, 1:4, 1:5
Surge Program
: Yes
Amplitude
: 0 - 100 mA
High Voltage (Twin Pulse)
Frequency
: 1- 200 Hz in steps of 1 Hz
Frequency Modulation (spectrum)
: 0 – 180 Hz in steps of 1 Hz, sum of Pulse frequency and
Frequency Modulation can not exceed 200 Hz
Modulation Program
: 1/1, 6/6, 12/12, 1/30/1/30 s
Surge Program
: Yes
Polarity
: Positive or Negative
Amplitude
: 0 – 500 Volt in steps of 1V
High Voltage Alternating (Twin Pulse)
Frequency
: 1 – 200 Hz in steps of 1 Hz
Frequency Modulation (spectrum)
: 0 – 180 Hz in steps of 1 Hz, sum of Pulse frequency and
Frequency Modulation can not exceed 200 Hz
Modulation program
: 1/1, 6/6, 12/12, 1/30/1/30 s
Alternating Period
: 10 -100 seconds in steps of 10
Ramp-in, ramp out
: 0,5 seconds
Amplitude
: 0 – 500 Volt in steps of 1 Volt
Micro Current
Frequency
: 0 – 1000 Hz
Polarity
: Positive or Negative
Surge Program
: Yes
Amplitude
: 10 µA - 1 mA in steps of 10 µA
Micro Current Alternating
Frequency
: 0 – 1000 Hz
Alternating Period
: 0,2 – 20 seconds, 0,2 – 1 s in steps of 0,1s, 1 – 20 s in steps of 1
s.
Ramp-in, ramp out
: 0 seconds
Amplitude
: 10 µA - 1 mA in steps of 10 µA
Diadynamic current
Settings
: MF, DF, CP, LP en CPid