24
!
CAUTION:
•
Do not place the device in a location where the power cord could be tripped over or
pulled out during treatment.
•
Do not attempt to use the device if it is not properly grounded. Make certain that the
device is electrically grounded by connecting it only to a grounded electrical service
receptacle conformable with the applicable national and local electrical codes regarding
medical environments.
•
Set power line switch [1] On (1)
•
Power LED indicator [5] is lit green indicating that the device is connected to the mains supply.
•
Turn on the device with push button [4]
•
The device will initialize and perform a self test. This may take a while.
•
At the end of the self test the device enters the Home menu and is ready for use.
6.4
Disconnection from mains supply
Systems without a battery:
•
When you have finished treatments turn the device off by setting the power line switch [1] to Off
(0). The device is now disconnected from the mains supply.
Systems with a battery:
•
Turn off the device with push button [4]
•
Power indicator LED [5] is still lit green, indicating that the device is still connected to the mains
supply and that the battery is being charged.
•
Set power line switch [1] Off (0) to stop charging and to disconnect the unit from the mains
supply.
6.5
Placing the optional battery
•
Remove the mains cable from the power line connector [1].
•
Place the 4-Series device upside down and on a soft surface.
•
Remove the two screws from the battery cover using the supplied screwdriver.
•
Slide and lift the battery cover.
•
Align the battery on the bottom of the main unit with the polarity of battery terminals in the correct
position. The polarity is marked at the bottom of the battery compartment.
•
Locate the black wire and attach it to the – terminal of the battery.
!
CAUTION:
•
Do not interchange the black and red wires as this will damage your device.
•
The battery contains material that is noxious to the environment. Observe the local
regulations when disposing of the battery. See also chapter 9.5.
•
Due to the high current demand of ultrasound applications, we recommend to explicitly
use batteries supplied by Enraf-Nonius B.V. (part number 2501016).
•
Locate the red wire and attach it to the + terminal of the battery.
•
Slide the battery upside down into the battery compartment taking care that the wires do not get
jammed.
•
Place and slide the battery cover back into position.
•
Secure the battery cover with the two screws using the supplied screwdriver.
•
Place the device back on its feet.
•
Reconnect the mains cable to the power line connector [1].
6.6
Operation from battery
•
Leave power line switch [1] in the Off position (0) and turn on the device on using push button [4].
•
Power LED indicator is lit orange, indicating that the device is operating from the battery.
•
The charge status of the battery is indicated in the right hand top corner of the display.