38
•
Enter the name of your favorite using the keyboard.
•
Touch
to store your favorite under the name just entered.
Notes:
•
Once saved, favorites can be retrieved from the
Electrotherapy, Ultrasound Therapy and Combination
Therapy menus.
•
4-polar treatments are automatically saved and loaded as a
dual channel treatment.
•
Vacuum settings are not saved.
8.2.3.6 Programming a Sequential Protocol
A sequential protocol consists of a series of treatment steps that are
executed in a sequence until the end of the protocol is reached.
•
Select Programming in the electrotherapy menu
•
When a treatment screen is completely set as required, touch the
“next” button to add a next step.
•
Continue with the remaining treatment steps until you have
reached the end of your protocol.
•
Touch the Store button in the navigation bar.
•
Enter the name of your sequential protocol as described in
paragraph 8.2.3.5.
Sequential protocols are stored under the Favorite menu.
8.2.3.7 System Settings
The Home menu gives access to all functions of the unit. Select in
the Home menu System Settings by touching the button “System
Settings”
The next screen appears