background image

EX-5175-EC

 

 

ENMET

 Corporation

 

10 

 
 

Normal Display Mode 

 

 

 

 

 

 

 

5ppm 

M

ENU

 

H

OLD

 the magnet over the 

M

ENU

 switch for 2 – 4 seconds 

to enter the Maintenance Menus 

Exit 

S

ELECT

 

Tap the 

S

ELECT

 switch to return to the Normal 

Display Mode.  

See Section 5.2.3

 

M

ENU

 

Tap the magnet over the 

M

ENU

 switch to cycle through Maintenance Menus

OR 

Zero 

Cal  OK 

If

the Zero signal is within Preset Specs the EX-5175-EC will 

display Cal OK, 

See Section 5.2.1 

If

 the Zero signal is not within Preset Specs the EX-5175-EC will 

display Bad ZERO  

N

OTE

:

 Some software revisions require the 

S

ELECT

 

switch be 

tapped to accept the signal.

 

S

ELECT

 

M

ENU

 

Bad ZERO 

PV: 0 

S

ELECT

Tap the 

S

ELECT

 switch to initiate Zero adjustment 

Apply

 Calibration Gas until signal value becomes 

stable (about 1 to 4 minutes) 

See 

Figure

 4

 

 

If

 cal is good display will indicate OK or Same 

 

If

 cal is not within preset “range” display will 

indicate Bad Sens 

 

M

ENU

 

PV: 0 

 20 

Cal  OK 

Same  mV 

Bad  Sens

Span 

S

ELECT

 

S

ELECT

 

OR 

OR 

N

OTE

: You can change the Calibration Gas Level. 

H

OLD

 the magnet over the 

S

ELECT

 switch  

The 

M

ENU

 switch changes digit indicated by underscore cursor

 

The 

S

ELECT

 switch locks underscored digit and moves to next digit

 

See Section 5.2.2

 

S

ELECT

mA Span 

M

ENU

 

S

ELECT

 

  100 

To change mA Span set point: 
Tap the 

M

ENU

 switch until mA Span is displayed 

Tap the 

S

ELECT

 switch to display the set point 

The 

M

ENU

 switch changes digit indicated by underscore cursor 

The 

S

ELECT

 switch locks underscored digit and moves to next digit 

To

 return to Normal Gas Display: 

Tap 

M

ENU

 switch until EXIT is displayed 

Then tap 

S

ELECT

 switch 

Select

Menu

To change Alarm set points: 
Tap Menu switch until Alarm to be changed is displayed 
Tap Select switch to display the set point 

The 

M

ENU

 switch: changes digit indicated by underscore 

cursor 
The 

S

ELECT

 switch: locks in the underscored digit and 

moves to next digit 

If change is not within range display returns to first digit 
If change is within range display moves to Set Time Delay 

Use 

M

ENU

 and 

S

ELECT

 switches as above to change time 

delay.  Between 0 and 5 seconds is allowed 

If change is within range display moves to next menu 
 

Λ

 - Indicates increasing alarm  

V

 - Indicates decreasing alarm  

SetTDsec

Alarm3 

S

ELECT

 

Λ

 50 

M

ENU

 

SetTDsec

Alarm1 

M

ENU

 

S

ELECT

 

Λ

 10 

SetTDsec

Alarm2 

M

ENU

 

S

ELECT

 

Λ

 20 

 

Figure 5: EX-5175-EC Maintenance Menu Flow chart 

Содержание EX-5175-EC

Страница 1: ...ENMET Corporation PO Box 979 Ann Arbor MI 48106 0979 www enmet com Manual Part Number 80003 099 MCN 13 008 11 06 13 EX 5175 EC Sensor Transmitter Manual...

Страница 2: ......

Страница 3: ...acement 11 6 0 REPLACEMENT PART NUMBERS 12 7 0 WARRANTY 12 List of Figures and Tables Figure 1 EX 5175 EC Features 2 Figure 2 EX 5175 EC Mounting Dimensions 3 Figure 3 Terminal Positions EX 5175 EC Se...

Страница 4: ......

Страница 5: ...and the contents After the inspection and after you have received written acknowledgment of the damage from the carrier contact ENMET Customer Service for return authorization and further instruction...

Страница 6: ...ment maintenance functions The switch locations are indicated by MENU and SELECT A magnetic field pulse is applied by momentarily putting the end of the magnet in proximity to the switch and then remo...

Страница 7: ...oosing locations of sensor transmitters Also take into consideration environmental factors when deciding on S T location Avoid locations where the S T may be damaged by liquid immersion excessive heat...

Страница 8: ...ct the wires from the controller power supply to the supplied J4 plug then attach to J4 terminal See Figure 3 for locations J4 PLUG TERMINAL TO CONTROLLER WIRING Position Function 1 24 VDC power 2 GND...

Страница 9: ...hen power is supplied to the EX 5175 EC the S T will display the following sequence of information NOTE Software revision may cause variations of display output Example of Display Function The instrum...

Страница 10: ...on below Display Measurement of the target gas Tap the magnet over the MENU button Display indicates Alarm 1 Set point Tap the magnet over the MENU button Display indicates Alarm 2 Set point Tap the m...

Страница 11: ...of target gas Hold the magnet over MENU switch for 2 5 seconds to enter the Maintenance Menu The Power Fault LED will flash Green Red to indicate the EX 5175 EC is in Maintenance Mode To exit the main...

Страница 12: ...C transmitters Calibration equipment is available from ENMET Corporation to calibrate the EX 5175 EC sensor transmitters A calibration adapter will have a fitting for the gas cylinder on one side and...

Страница 13: ...Figure 4 3 Open the valve to apply the calibration gas to the sensor wait approximately 120 seconds 2 minutes 4 Tap the SELECT switch to perform a Span procedure The display will alternate between th...

Страница 14: ...anges digit indicated by underscore cursor The SELECT switch locks underscored digit and moves to next digit See Section 5 2 2 SELECT mA Span MENU SELECT 100 To change mA Span set point Tap the MENU s...

Страница 15: ...d Assembly into the sensor housing cover Reassemble the sensor housing Turn On the sensor transmitter After the new sensor has been installed it is suggested to allow the sensor to stabilize for 24 ho...

Страница 16: ...edEx Ground If an expedient means of transportation is requested during the warranty period the customer is responsible for the difference between the most economical means and the expedient mode ENME...

Страница 17: ...processing delays ENMET charges a one hour minimum billing for all approved repairs with additional time billed to the closest tenth of an hour All instruments sent to ENMET are subject to a minimum...

Страница 18: ......

Страница 19: ..._____________ Payment Terms COD Check one VISA MasterCard______________________ ________ ________ Card number Expiration Card Code American Express______________________ ________ ________ Card number...

Страница 20: ......

Отзывы: