5.0 Maintenance of the EX-5175-EC 

C

AUTION

:

 

Do not open the 

EX-5175-EC 

S/T 

in a classified area.

 

C

AUTION

:

 

Do Not Attempt A Span Procedure Without Calibration Gas Applied to The Sensor

; if this is done, the S/T is 

forced into a calibration fault mode. 

Magnetic switches control the

 

M

ENU

 and 

S

ELECT 

functions.  The 

M

ENU

 and 

S

ELECT

 switch locations are indicated on 

the display panel, see 

Figure 3.

  The 

M

ENU

 switch is used to display the various menu options and make incremental 

changes to numbers such as alarm points, calibrations gas, etc.  The 

S

ELECT

 switch is used to select that option, set 

zero or span digit.  Most maintenance functions are controlled by simple taps of the supplied magnet on the transmitter 
glass, below the 

M

ENU

 

and 

S

ELECT

 

boxes on the front panel. 

5.1 Maintenance Menu 

To enter the maintenance menu hold the magnet over the 

M

ENU

 switch for 2 to 4 seconds 

Table 1

 indicates the maintenance menu sequence see 

Figure 5

 for a detailed maintenance menu flow chart. 

Table 1: EX-5175-EC maintenance Menus Sequence 

Example of Display 

Function 

 
 

 

Normal Display Mode 

Measurement of target gas 

Hold the magnet over 

M

ENU

 switch for 2 – 5 seconds to enter the Maintenance Menu 

The Power/Fault LED will flash Green – Red to indicate the 

EX-5175-EC

 is in Maintenance Mode 

 

To exit the maintenance Menu and return to the Normal 
Display Mode: 

If intended function Tap the magnet over 

S

ELECT

 switch 

Tap the magnet over the 

M

ENU

 switch to advance to the Zero procedure 

 

For adjusting Zero: 

If intended function Tap the magnet over 

S

ELECT

 switch 

N

OTE

:

 Not applicable for Oxygen S/T. 

Tap the magnet over the 

M

ENU

 switch to advance to the Span procedure 

 

For adjusting the Span: 
If intended function Tap the magnet over 

S

ELECT

 switch 

 

Tap the magnet over the 

M

ENU

 switch to advance to each Alarm set point procedures 

 

For adjusting the Alarm 1, 2 and 3 set points: 
If Intended function Tap the magnet over 

S

ELECT

 switch 

 

Tap the magnet over the 

M

ENU

 switch to advance the mA Span set point procedure 

 

For adjusting the mA Span set point: 
If intended function Tap the magnet over 

S

ELECT

 switch 

 

 

Taping the 

M

ENU

 switch without taping the 

S

ELECT

 switch will allow you to cycle through the menu options. 

You must Tap the 

S

ELECT

 switch in order to change the desired operation. 

N

OTE

:

 If the S/T fails to respond, the magnet may have become weak and may need to be replaced. 

5ppm 

Exit 

Zero 

Span 

mA Span 

Alarm1 

Alarm2 

Alarm3 

Содержание EX-5175-EC

Страница 1: ...ENMET Corporation PO Box 979 Ann Arbor MI 48106 0979 www enmet com Manual Part Number 80003 099 MCN 13 008 11 06 13 EX 5175 EC Sensor Transmitter Manual...

Страница 2: ......

Страница 3: ...acement 11 6 0 REPLACEMENT PART NUMBERS 12 7 0 WARRANTY 12 List of Figures and Tables Figure 1 EX 5175 EC Features 2 Figure 2 EX 5175 EC Mounting Dimensions 3 Figure 3 Terminal Positions EX 5175 EC Se...

Страница 4: ......

Страница 5: ...and the contents After the inspection and after you have received written acknowledgment of the damage from the carrier contact ENMET Customer Service for return authorization and further instruction...

Страница 6: ...ment maintenance functions The switch locations are indicated by MENU and SELECT A magnetic field pulse is applied by momentarily putting the end of the magnet in proximity to the switch and then remo...

Страница 7: ...oosing locations of sensor transmitters Also take into consideration environmental factors when deciding on S T location Avoid locations where the S T may be damaged by liquid immersion excessive heat...

Страница 8: ...ct the wires from the controller power supply to the supplied J4 plug then attach to J4 terminal See Figure 3 for locations J4 PLUG TERMINAL TO CONTROLLER WIRING Position Function 1 24 VDC power 2 GND...

Страница 9: ...hen power is supplied to the EX 5175 EC the S T will display the following sequence of information NOTE Software revision may cause variations of display output Example of Display Function The instrum...

Страница 10: ...on below Display Measurement of the target gas Tap the magnet over the MENU button Display indicates Alarm 1 Set point Tap the magnet over the MENU button Display indicates Alarm 2 Set point Tap the m...

Страница 11: ...of target gas Hold the magnet over MENU switch for 2 5 seconds to enter the Maintenance Menu The Power Fault LED will flash Green Red to indicate the EX 5175 EC is in Maintenance Mode To exit the main...

Страница 12: ...C transmitters Calibration equipment is available from ENMET Corporation to calibrate the EX 5175 EC sensor transmitters A calibration adapter will have a fitting for the gas cylinder on one side and...

Страница 13: ...Figure 4 3 Open the valve to apply the calibration gas to the sensor wait approximately 120 seconds 2 minutes 4 Tap the SELECT switch to perform a Span procedure The display will alternate between th...

Страница 14: ...anges digit indicated by underscore cursor The SELECT switch locks underscored digit and moves to next digit See Section 5 2 2 SELECT mA Span MENU SELECT 100 To change mA Span set point Tap the MENU s...

Страница 15: ...d Assembly into the sensor housing cover Reassemble the sensor housing Turn On the sensor transmitter After the new sensor has been installed it is suggested to allow the sensor to stabilize for 24 ho...

Страница 16: ...edEx Ground If an expedient means of transportation is requested during the warranty period the customer is responsible for the difference between the most economical means and the expedient mode ENME...

Страница 17: ...processing delays ENMET charges a one hour minimum billing for all approved repairs with additional time billed to the closest tenth of an hour All instruments sent to ENMET are subject to a minimum...

Страница 18: ......

Страница 19: ..._____________ Payment Terms COD Check one VISA MasterCard______________________ ________ ________ Card number Expiration Card Code American Express______________________ ________ ________ Card number...

Страница 20: ......

Отзывы: