EX-5175-EC

 

 

ENMET

 Corporation

 

 
 

5.2.1 Zero Adjust 

Zero function is Not applicable for Oxygen sensor/transmitters, go to Section 5.2.2 Gas Span. 

A ZERO function should be performed only when the 

EX-5175-EC 

sensor/transmitter is exposed to fresh air.  If the air 

at the sensor is in question, use a cylinder of 20.9% oxygen to provide a clean air reference.  Attach the cylinder to the 
calibration adapter. 

Enter the maintenance menu by placing the magnet over 

M

ENU

 

switch for 2 to 4 seconds.  See 

Figure 5,

 

EX-5175-EC 

Maintenance Menu flow chart. 

The second menu available is the Zero. 

Tap the 

S

ELECT

 switch to perform a Zero. 

 

If the Zero is successful

, Cal OK appears on the display and in 1 – 2 seconds, display will change to Span. 

If you wish to Span the sensor Tap the 

S

ELECT

 

switch you are now ready to apply gas.  

Proceed to gas span step 2

 

If you wish to Exit the maintenance menu, Tap 

M

ENU

 

switch until Exit is displayed, then tap 

S

ELECT

 

switch to 

return to the instrument Normal Gas Display 

 

If the

 

Zero is Not successful,

 sensor is outside of safe parameters to be zeroed, the display will read Bad Zero.  

Repeat Section 5.2.1 Zero Adjust making sure to use a cylinder of 20.9% Oxygen. 

5.2.2 Gas Span 

It is recommended that the Zero Function be performed first.

 

Enter the maintenance menu.  See 

Figure 5,

 

EX-5175-EC 

Maintenance Menu flow chart. 

1. 

Tap the 

M

ENU

 

switch once to show Span on the display. 

2. 

Attach the associated calibration gas cylinder to the regulator and calibration cover.  See 

Figure

 

4

 

 

 

3. 

Open the valve to apply the calibration gas to the sensor, wait approximately 120 seconds(2 minutes) 

4. 

Tap the 

S

ELECT

 

switch to perform a Span procedure.  The display will alternate between the calibration gas 

concentration and a signal level. 

5. 

Watch for the signal level to stabilize.  Typical 2 – 4 minutes. 

6. 

Once the signal level has stabilized, the 

EX-5175-EC 

will automatically lock in the calibration data and: 

 

If the Span is successful, Cal OK appears on the display momentarily, then advances to Alarm 1.  Remove 
calibration gas.  To exit maintenance menu tap the 

M

ENU

 

switch until Exit appears, then tap the 

S

ELECT

 

switch. 

 

If the sensor is outside of acceptable parameters, Bad Span is displayed momentarily, then returns to Span. 
Remove calibration gas.  Tap the 

M

ENU

 

switch until Exit appears, then tap the 

S

ELECT

 

switch.  Check span gas 

and repeat calibration in 30 – 60 minutes. 

 

If the sensor did not respond to gas, Same mV is displayed momentarily, then returns to Span. 

 

Remove calibration gas, tap the 

M

ENU

 

switch until Exit appears, then tap the 

S

ELECT

 

switch and try 

calibration again in 30-60 minutes. 
If the sensor will not calibrate See Section 5.4. 

  N

OTE

:

 Some software revisions require the 

S

ELECT

 

switch be tapped to accept the signal. 

7. 

Calibration is complete. 

 

5.2.3 Exit Maintenance Menu 

Exit maintenance, by tapping on the 

M

ENU

 switch until Exit appears on the display.  Tap the 

S

ELECT

 switch to return 

to the instrument Normal Gas Display. 

N

OTE

You can change the Calibration Gas Level.  

H

OLD

 the magnet over the 

S

ELECT

 switch for 2 – 4 seconds

 

The 

M

ENU

 switch changes digit indicated by underscore cursor

 

The 

S

ELECT

 switch locks underscored digit and moves to next digit

 

Содержание EX-5175-EC

Страница 1: ...ENMET Corporation PO Box 979 Ann Arbor MI 48106 0979 www enmet com Manual Part Number 80003 099 MCN 13 008 11 06 13 EX 5175 EC Sensor Transmitter Manual...

Страница 2: ......

Страница 3: ...acement 11 6 0 REPLACEMENT PART NUMBERS 12 7 0 WARRANTY 12 List of Figures and Tables Figure 1 EX 5175 EC Features 2 Figure 2 EX 5175 EC Mounting Dimensions 3 Figure 3 Terminal Positions EX 5175 EC Se...

Страница 4: ......

Страница 5: ...and the contents After the inspection and after you have received written acknowledgment of the damage from the carrier contact ENMET Customer Service for return authorization and further instruction...

Страница 6: ...ment maintenance functions The switch locations are indicated by MENU and SELECT A magnetic field pulse is applied by momentarily putting the end of the magnet in proximity to the switch and then remo...

Страница 7: ...oosing locations of sensor transmitters Also take into consideration environmental factors when deciding on S T location Avoid locations where the S T may be damaged by liquid immersion excessive heat...

Страница 8: ...ct the wires from the controller power supply to the supplied J4 plug then attach to J4 terminal See Figure 3 for locations J4 PLUG TERMINAL TO CONTROLLER WIRING Position Function 1 24 VDC power 2 GND...

Страница 9: ...hen power is supplied to the EX 5175 EC the S T will display the following sequence of information NOTE Software revision may cause variations of display output Example of Display Function The instrum...

Страница 10: ...on below Display Measurement of the target gas Tap the magnet over the MENU button Display indicates Alarm 1 Set point Tap the magnet over the MENU button Display indicates Alarm 2 Set point Tap the m...

Страница 11: ...of target gas Hold the magnet over MENU switch for 2 5 seconds to enter the Maintenance Menu The Power Fault LED will flash Green Red to indicate the EX 5175 EC is in Maintenance Mode To exit the main...

Страница 12: ...C transmitters Calibration equipment is available from ENMET Corporation to calibrate the EX 5175 EC sensor transmitters A calibration adapter will have a fitting for the gas cylinder on one side and...

Страница 13: ...Figure 4 3 Open the valve to apply the calibration gas to the sensor wait approximately 120 seconds 2 minutes 4 Tap the SELECT switch to perform a Span procedure The display will alternate between th...

Страница 14: ...anges digit indicated by underscore cursor The SELECT switch locks underscored digit and moves to next digit See Section 5 2 2 SELECT mA Span MENU SELECT 100 To change mA Span set point Tap the MENU s...

Страница 15: ...d Assembly into the sensor housing cover Reassemble the sensor housing Turn On the sensor transmitter After the new sensor has been installed it is suggested to allow the sensor to stabilize for 24 ho...

Страница 16: ...edEx Ground If an expedient means of transportation is requested during the warranty period the customer is responsible for the difference between the most economical means and the expedient mode ENME...

Страница 17: ...processing delays ENMET charges a one hour minimum billing for all approved repairs with additional time billed to the closest tenth of an hour All instruments sent to ENMET are subject to a minimum...

Страница 18: ......

Страница 19: ..._____________ Payment Terms COD Check one VISA MasterCard______________________ ________ ________ Card number Expiration Card Code American Express______________________ ________ ________ Card number...

Страница 20: ......

Отзывы: