AirGuard
TM
25
www.bullard.com
Connect the regulator to the calibration port on the side of the AirGuard.
Open the gas regulator fully by turning the on/off knob two turns counter clockwise.
Upon selecting the “NEXT”
t
he AirGuard will display a “Zero in Progress” display with a timer which counts down
the time remaining in the zero function.
Upon completion, the AirGuard will display a screen indicating the sensor voltage (mV) and also report sensor flow,
main flow, and results. A green check indicates a successful event.
Sensor voltage (mv), Zero gas reading can vary between 25 mV to 100mV.
Calibration
Connection
Charger
Connection
Remote
Alarm