Bump Test
It is recommended that the AirGuard is bump checked before use to ensure proper response to a known
calibration standard. Should the reported test values from the bump check not meet the calibration standard
value, ±10% of the reference value, it should be recalibrated.
BULLARD provides a calibration kit for the AirGuard with all the components to complete this test.
(Part Number: AGDCK)
Warning!
To perform a bump check, the AirGuard must be disconnected from the
compressor air supply.
Calibration cylinder must be connected to a pre-determined or fixed flow regulator assembly, with a male quick
connect fitting and tubing attached to the regulator.
The calibration cylinder assembly tubing is then attached to the female calibration connection located on the right-
hand side of the instrument and the cylinder is turned on to initiate gas flow.