VER. A 09 / 24 / 2009

Page 1 of 31

Содержание Son-Mate II

Страница 1: ...VER A 09 24 2009 Page 1 of 31 ...

Страница 2: ...speed veterinary dental air unit with ultrasonic scaler Low speed high speed air water syringe Excelsior high speed veterinary dental air unit with vacuum Electrosurgery ultrasonic scaler Low speed high speed air water syringe Tri Mate scaler polisher electro surge for veterinary use only and A D S 2000 Anesthesia delivery system Ventilator for veterinary use only We manufacture all of the inserts...

Страница 3: ...ltrasonic scalers should not be used by or in close proximity to anyone who has any electronic implant device pacemaker defibrillator or has any neurological problems PLEASE READ VERY CAREFULLY Engler Engineering Corporation makes every effort to verify that all parts for the device along with any optional accessories ordered were shipped from our facility in Hialeah Florida and are included in th...

Страница 4: ...TION 16 POLISHER MAINTENANCE PROPHY ANGLE 16 STRAIGHT HANDPIECE MAINTENANCE 17 MICROMOTOR MAINTENANCE 18 SCALER TROUBLESHOOTING 19 POLISHER TROUBLESHOOTING 20 WATER FILTER CLEANING INSTRUCTIONS 21 SADDLE VALVE ASSEMBLY 22 WATER TANK INSTRUCTIONS 25 CARE AND MAINTENANCE OF YOUR PORTABLE WATER TANK 26 OPTIONAL SON MATE II ACCESSORIES 27 WARRANTY INFORMATION 29 RETURN FOR EVALUATION REPAIR FORM 30 SO...

Страница 5: ... 60 PSI max III POLISHER Power Control Variable Voltage Style Output Voltage 24 VDC Output Current 5 Amperes max IV MICROMOTOR Velocity Range 300 30 000 RPM Style Sealed Bearing E type Brush Design Fully Replaceable V STRAIGHT HANDPIECE Maximum Drive Velocity 30 000 RPM Style Sealed Bearing E type VI PROPHY ANGLE Recommended Drive Velocity 3 000 5 000 RPM Type Sealed Bearing Open Bushing IMPORTANT...

Страница 6: ... CHASSIS DIMMESION Length 9 in 23 cm Width 8 in 20 cm Height 3 3 in 8 4 cm X CABLE LENGTHS Handpiece Footswitch Power Cord Water Line 96 in 244 cm 96 in 244 cm 72 in 183 cm 96 in 244 cm XI WEIGHT Net Shipping 3 Lbs 1 36 Kg 7 Lbs 3 17 Kg Page 6 of 31 ...

Страница 7: ...k Disconnect so that the water can be completely shut off and line pressure relieved when the unit is not in use This device comes equipped with an 8 foot 244 cm water line a male Poly Quick disconnect fitting a water filter and a coupling body The water line must be connected to the white fitting located on the back of the device before a connection is made to the water source To connect the wate...

Страница 8: ...I Use this to create a custom water installation 44300 Female quick disconnect 2 Saddle Valve Assembly P N A44303 This kit contains all parts to quickly and easily connects your unit to an existing cold water supply line 3 Faucet Adapter Assembly P N A22303 This screws on the faucet and has a female mating connector Page 8 of 31 ...

Страница 9: ...and inspected quarterly and we suggest that the filter disc and O ring be changed out at least once a year See page 17 for instructions CONNECTING THE FOOT SWITCH To connect the footswitch to the unit insert the male quick disconnect into the female quick disconnect on the rear of the device and then tighten the securing nut by turning it clockwise CONNECTING POWER SUPPLY First plug the power cord...

Страница 10: ...water control to a level where you have a fine mist at the tip Higher power settings will result in hotter water NOTE Tips sent from our facility are not pre sterilized IMPORTANT If you find that tightening the nosecone by hand is not successful you may lightly tighten the nosecone with the tip wrench Since the tip wrench is designed to remove the tips it is important that you DO NOT over tighten ...

Страница 11: ...s to form and fall from the tip when unit is disengaged IMPORTANT Pressure on the tip is not necessary to remove calculus or tartar Enamel on the teeth may be damaged or removed when using excessive pressure It can also be damaged if the scaling tip is left to rest in one spot for even a few seconds Using the tip as a pry to remove calculus or tartar may change the shape of the tip which in turn c...

Страница 12: ...is procedure could be done by spraying a fine mist of sterilization solution on the unit allowing it to remain on the chassis for the length of time recommended by the manufacturer The surface should be wiped with a clean damp cloth or as suggested by the Chemiclave manufacturer Dry completely IMPORTANT In using any chemical sterilization solution please follow the manufacturer s suggested procedu...

Страница 13: ...or handpiece TO INSTALL THE STACK Carefully remove the new stack from the clear shipping tube A Locate the hole see photo above on the side of the stack Align it with the white dot on the top of the handpiece and let the stack slide down into the handpiece B Gently push the stack down until it stops There should be approximately inch showing above the handpiece DO NOT force the stack into the Hand...

Страница 14: ...Page 14 of 31 ...

Страница 15: ...ow to mid range speed setting for polishing High speed will damage the gears in the Prophy angle SAFETY INFORMATION Never turn the Lock Ring while the handpiece is in operation 1 Never reverse the direction of the micromotor while it is in operation Possible damage to the unit may occur 2 Do not drop the Micromotor Handpiece or Prophy Angle on the floor or other hard surface as the shock may cause...

Страница 16: ...tooth with a circular motion Do not allow the rubber cup to remain stationary on one area for an extended period of time POLISHER MAINTENANCE PROPHY ANGLE The Prophy Angle is a precision engineered dental device All of the gears and shaft assemblies are made of high grade stainless steel which if cleaned and lubricate correctly will provide long trouble free service Daily Cleaning and Lubrication ...

Страница 17: ... Type nozzle into the bottom of the handpiece Holding the two together tightly with can in the upright position push spray button for 2 to 3 seconds NOTE If spray time is too short oil may not be propelled into all areas of the handpiece CLEANING and STERILIZATION OF HANDPIECE Cleaning 1 Wipe the handpiece clean with an alcohol soaked soft tissue 2 Never clean the handpiece with boiling water chem...

Страница 18: ...rranty MICROMOTOR MAINTENANCE The Micromotor is capable of rotational speeds up to 35 000 RPM for use in cutting sectioning and drilling It contains sealed bearings and does not require any lubrication The Micromotor has cooling vents located at the back of its body DO NOT allow water oil or any other substance to enter these vents Failure to keep debris out of the Micromotor will shorten the life...

Страница 19: ... INDICATOR LIGHTS UP LITTLE OR NO VIBRATION CAVITATION ON TIP A Tip loose tighten the tip B Tip damaged replace the tip C Old or damaged stack replace the stack D Contact Engler Engineering Corporation HOT WATER AND OR SCALING HANDPIECE The Stack requires a constant cool water flow in order to maintain tip water temperature below 100 degrees F You may correct the problem by A Adjusting water flow ...

Страница 20: ...ISHER TROUBLESHOOTING HOT MICROMOTOR I Straight handpiece causing drag lubricate correctly II Worn brushes in micromotor Contact Engler Engineering Corporation III Oil inside micromotor return to Engler Engineering Corporation INTERMITTENT OPERATION I Unit polishes and then stops a Damaged micromotor cord contact Engler Engineering Corporation b Damaged footswitch contact Engler Engineering Corpor...

Страница 21: ...th an outer male thread and another with an inner female thread Take the side with the inner female thread and look into it should see an O RING and under it a FILTER DISC Refer to Figure below 4 Remove the O RING by carefully prying it out of the inner wall with a fingernail or other dull object NOTE Be careful not to rip or break this O RING a water leak could result 1 Next remove the FILTER DIS...

Страница 22: ...saddle valve over and around your existing 3 8 copper tubing Evenly tighten the securing screws so that the saddle valve is securely fastened onto the copper tubing DO NOT OVERTIGHTEN THIS MAY CAUSE THE COPPER TUBING TO CRIMP AND REDUCE FLOW 2 To Mount the Female Quick Disconnect YOU HAVE TWO OPTIONS Page 22 of 31 ...

Страница 23: ...Page 23 of 31 ...

Страница 24: ...e Then proceed to place the hose coupling nut over the 1 4 water tubing Place this tubing into the back of the female quick disconnect and tighten the hose coupling nut securely 1 To Pierce Existing Water Tube You are now ready to pierce the copper tube Turn the handle of the saddle valve in a CLOCKWISE direction until it will go no further The Saddle valve is now installed and in the OFF position...

Страница 25: ...allow air to escape 2 Loosen cap slowly Remove pump cap assembly Pour out any remaining liquid rinse all parts thoroughly with clean water 3 Always store tank empty and with tank cap loose upside down TROUBLESHOOTING PROBLEM TANK FAILS TO PRESSURIZE 1 Be sure cap is tight 2 Check to see if pressure relief valve is in safety position 3 Remove the pump from the tank Turn pump handle counterclockwise...

Страница 26: ...ief valve hole as this could cause the tank to explode DO NOT pressurize the tank until ready for use DO NOT lift or carry the tank by waterline extension rod or pump handle unless it is securely locked in place CARE AND MAINTENANCE OF YOUR PORTABLE WATER TANK To keep slime from forming inside the tank and eventually getting into the dental unit Every two weeks dispose of water in tank Pour 1 2 cu...

Страница 27: ...INGS A Do not attempt to disassemble B DO NOT SUBMERGE IN LIQUIDS INCLUDING ULTRASONIC SOLUTIONS C Do not heat over 275 ºF 135 ºC D Use only Engler Care Free Prophy Rubber Cups Other brands will not properly seal the angle causing premature wear and voiding the warranty E Use 1 year and dispose of properly YOUR CARE FREE ANGLE IS WARRANTED AGAINST DEFECTS IN MATERIAL AND WORKMANSHIP FOR 6 MONTHS A...

Страница 28: ...concerns please contact us ENGLER ENGINEERING CORP 800 445 8581 To contact us by internet please use info englerusa com To help troubleshoot problems download manuals and see our products Visit us at www engler411 com Page 28 of 31 ...

Страница 29: ...thirty 30 days from date of purchase Prophy warranted for a period of thirty 30 days from date of purchase All warranties are void if the items have been damaged due to negligence improper use failure to lubricate removed parts or opening the control unit PLEASE NOTE Whenever returning any products for evaluation and or repair we strongly suggest you send in the complete unit with all attachments ...

Страница 30: ...__ State ___________________ Zip _______________________________ ITEM BEING SENT _________________________________ SERIAL _______________________________ Please describe what is happening or why you are sending in this unit When returning items to us we would prefer you send the complete unit so that a proper evaluation can be made It is suggested that you ship all returns to us by Federal Express...

Страница 31: ... 800 445 8581 on the day the loaner is received if any items are missing 8 All loaner items must be returned in good working condition We include a copy of our check list for you to use to verify all parts are being returned Missing and or damaged items will be charged to your credit card NOTE Loaner requests received after 11 00 AM Eastern time will be shipped out the next business day Unless spe...