5. Autoclave the Handpiece for no longer than 20 minutes at 121 C (250 F), or 15 min. at 132 C
(270 F).
Keep the straight Handpiece away from water vapor or mist that may settle and cause premature
damage to the bearings.
IMPORTANT:
If you experience problems during operation, call our repair department. DO NOT
attempt to repair the straight Handpiece. Doing so may shift the internal springs, causing
permanent damage to the unit and will void warranty.
MICROMOTOR MAINTENANCE:
The Micromotor is capable of rotational speeds up to 35,000 RPM for use in cutting, sectioning and
drilling. It contains sealed bearings and does not require any lubrication. The Micromotor has
cooling vents located at the back of its body,
DO NOT
allow water, oil, or any other substance to
enter these vents. Failure to keep debris out of the Micromotor will shorten the lifespan of the unit
and cause permanent damage.
IMPORTANT:
1. NEVER change the direction of the Micromotor while it is in operation. ALWAYS wait until
it has come to a full stop.
2. NEVER oil, or allow oil to get into the micromotor.