EQ-1500 Operation Manual Rev 6
19
C h a p t e r 3
INSTALLATION
Installation of the EQ-1500 consists of connecting electrical, water, and gas supplies, and
connecting the lamp output to the user’s equipment.
Unpacking
Upon arrival, start by inspecting all parts of the system for completeness and any damage
incurred in shipping. The EQ-1500 shipping box should contain: 1) EQ-1500 Controller unit,
1) EQ-1500 Lamp House, 1) Lamp House Signal/Power Cable, and 1) 2m IEC power cord.
If any part is missing or appears damaged, contact the manufacturer. Do not attempt to
substitute any parts. There are no user-serviceable parts inside the EQ-1500 Lamp House or
Controller unit.
The EQ-1500 system is shipped with the laser fiber optic cable already connected between the
Controller and Lamp House. Use care during installation and operation to avoid creating
sharp bends or kinks in the cable. Minimum bend radius for the cable is 60 mm (2.4 in).
Facilities Connections
Electrical Service
The EQ-1500 can be operated from 100–120/200–240VAC, single-phase, 50-60Hz. Power
consumption is approximately 200VA during normal operation, 300VA during lamp starting.
The EQ-1500 uses a detachable three-wire standard IEC 60320 C13 cord for connection to
the power source. The cord also provides the connection to the protective earth circuit.
Exposed metal parts of the EQ-1500 are connected to the protective earth ground via the cord
and the outlet ground, providing protection against electrical shock. Never operate the EQ-
1500 from an outlet without a properly connected protective earth ground.
There is no selection switch for line voltage, it is automatically adjusted internally.
Содержание LDLS EQ-1500
Страница 4: ...iv...
Страница 6: ...vi...
Страница 14: ...EQ 1500 Operation Manual Rev 6 8...
Страница 24: ...EQ 1500 Operation Manual Rev 6 18...
Страница 28: ...EQ 1500 Operation Manual Rev 6 22 Plasma Location Figure 8 Optical interface...
Страница 36: ...EQ 1500 Operation Manual Rev 6 30...
Страница 42: ...EQ 1500 Operation Manual Rev 6 36...
Страница 43: ......