Installation
Turbimax CUS71D
12
Hauser
2. Mount the sensor so that it is straight and parallel to the basin wall (measured value
offset).
Do not install the sensor in zones where air bubbles, turbulence, high
concentrations of turbid material and suspended matter, or foam formation occur
(e.g. inlet).
3. Using a dip pipe, install the sensor at least 20 cm (0.66 ft) beneath the surface of the
water.
5.1.2
Dimensions
Ø 74 (2.9)
Ø 75 (3.0)
Ø 33.7 (1.3)
20.5
(0.8)
1
7
(0
.7
)
1
5
(0
.6
)
G1"
NPT 3/4"
43 (1.7)
68 (2.7)
60.5 (2.4)
AF 27
Ø 34.7 (1.4)
A0036897
2
Dimensions for standard sensor.
Dimensions: mm (in)