Technical data
TOC Analyzer CA79
50
Hauser
14
Technical data
14.1 Input
Measured variable
TOC
Measuring range
0.5 to 1
000 µg/l (ppb)
Input signal
Controller input 24 V (order option)
The controller input starts a measurement. The function is only available for 1-channel
devices.
14.2 Output
Output signal
Measuring channel 1
0/4 to 20 mA, galvanically isolated
Measuring channel 2 (optional)
0/4 to 20 mA, galvanically isolated
Signal on alarm
1 port for waste UNF ¼ - 28
Load
Max. 500 Ω
Transmission behavior
Adjustable, in the measuring range 4 to 20 mA
Standby: 3.8 mA
14.3 Current outputs, active
Span
0 to 20 mA; according to Namur NE43