TOC Analyzer CA79
Commissioning
Hauser
21
A0046944
8
Leak in the system
UV intensity too low
If the intensity of the UV lamp is too low or if the UV lamp is defective, the analyzer stops
the measurement automatically and the
UV lamp broken
message is displayed. In this
case, a new UV reactor is needed.
L
WARNING
Radiation source with short-wave UV radiation!
Improper handling can cause damage to eyes and skin!
‣
Before working on the reactor, always take the device out of service and disconnect it
from the power supply!
‣
Always replace the reactor as a complete assembly!
‣
Take damaged reactors out of service!
‣
Never open the reactor to replace individual components!
‣
Ensure the insulation at the reactor ends is intact (undamaged heat shrink tubes)!
A0046945
9
UV sensor error message