TOC Analyzer CA79
Product description
Hauser
11
4.3.2
Socket assignment (system with two connection sockets)
Output 1
Socket
Pin
Description
A0046897
1:
4 (0) - 20 mA (GND)
2:
4 (0) - 20 mA (+)
TOC 1 (0 to set limit)
3:
4 (0) - 20 mA (GND)
4:
4 (0) - 20 mA (GND)
Δ conductivity (0 to set limit)
5:
Relay
6:
Relay
Group error message or limit value overshoot depending on the
selected option
Pin 1/2:
Provides an analog signal between 4 and 20 mA for the TOC measured value at
the SAMPLE input between 0 and the limit (to be configured in the
Settings
menu,
Options 1
tab).
Pin 3/4:
Provides an analog signal between 4 and 20 mA for the Δ conductivity value
between 0 and the limit in μS/cm (to be configured in the
Settings
menu,
Service 1
tab).
Pin 5/6:
Acts as an NO contact if the configured TOC or conductivity limit value is
exceeded in the "analog output" option, or as an NC contact in the option with a group error
message in the event of a power failure, a reaction of the leak, lamp or pressure sensor
and if one of the two conductivity sensors is over range.
4.3.3
Output 2
Socket
Pin
Description
A0046897
1:
4 (0) - 20 mA (GND)
2:
4 (0) - 20 mA (+)
TOC 2 (0 to set limit)
3:
4 (0) - 20 mA (GND)
4:
4 (0) - 20 mA (GND)
TOC 3 (0 to set limit)
5:
Controller (GND)
6:
Controller (+)
Controller input / trigger for 24 V DC
Pin 1/2:
Provides an analog signal between 4 and 20 mA for the TOC measured value at
INPUT 2 between 0 and the limit (to be configured in the
Settings
menu,
Options 1
tab).
Pin 3/4:
Provides an analog signal between 4 and 20 mA for the TOC measured value at
INPUT 3 between 0 and the limit (to be configured in the
Settings
menu,
Options 1
tab).
Pin 5/6:
External controller input / trigger, measurement is active when the voltage is
applied and stops if the voltage is 0 V.
The measured value is updated every minute. During a calibration or the SST, the last value
is displayed until a new measurement is started.
Содержание TOC Analyzer CA79