TDLAS Gas Analyzer
XA02744C
Hauser
35
Equipment
Installation
The information in this chapter is related to safety considerations for equip
-
ment installation.
Physical
Class I, Div 2
Enclosure type
Type 4X stainless steel enclosures
Dimensions
1074 mm H x 508 mm W x 279 mm D
(42.3 x 20 x 11 inches)
Weight
approximately
45 kg (100 lbs)
Sample cell
dimensions
438 mm H x 108 mm W
(17.3 x 4.3 inches)
Sample cell
construction
316L series polished stainless steel
Number of sample
cells
2
Area classification
Certification
Hauser Class I Division 2 analyzers use a non-
incendive protection method, and as such all portions of the
local installation codes apply. The maximum allowed
inductance to resistance ratio (L/R ratio) for the field wiring
interface must be less than 25 µH/Ω. The maximum total
loop capacitance shall be 0.27 microfarads.
The safety of the analyzer is the responsibility of the installer
and the organization he/she represents.
Configurations requiring optional accessories, e.g., probe
assemblies, with specific characteristics must meet
manufacturer specifications.