XA02744C
TDLAS Gas Analyzer
30
Hauser
Sample cell pressure
range
700 to 1400 mbara
700 to 1700 mbara —
optional
Sample cell
temperature range
-20 °C to 50 °C (-4 °F to 122 °F)
-10 °C to 60 °C (14 °F to 140 °F) —
optional
Maximum cell
pressure
70 kPag (10 psig)
Sample flow rate
0.5 to 1.0 slpm (1 to 2 scfh)
Bypass flow rate
1 slpm (2 scfh)
Electrical and communication
Input Voltage
(Electronics)
120 - 240 VAC ±10%, 50 to 60 Hz, 20W Max
24 VDC, 1.6A Maximum
12 VDC, 3.2 A Maximum
Input Voltage
SCS Heater
(Optional)
120 VAC or 240 VAC +/-10%, 50 to 60 Hz sin
-
gle phase, 200 W
Max current
(Unheated)
0.8 A maximum at 120 VAC
1.6 A at 24 VDC, 3.2 A at 12 VDC
Max current
(heated)
2 A maximum at 120 VAC
Communication
Analog: 2 4-20mA isolated, 1200 ohms at 24
VDC max load
Serial: RS232C —
standard
, RS485 and Ethernet
—
optional
Protocol: Modbus Gould RTU or Daniel RTU or
ASCII
Alarms
2, general fault and concentration alarms via
Modbus and analog output(s)
LCD display
Concentration, cell pressure and temperature,
diagnostics
Physical
Enclosure type
Type 4X - 304 stainless steel
Содержание SS2000