31
Table 5–2 SS2000e analyzer specifications
Dimensions
973 mm H x 406 mm W x 229 mm D
(38.3 x 16 x 9 inches)
Weight
approximately
34 kg (75 lbs)
Sample cell
dimensions
438 mm H x 108 mm W
(17.3 x 4.3 inches)
Number of sample
cells
1
Area classification
Certification
Measurement Data
Target Components
H
O or CO
in Natural Gas
Principle of
measurement
Tunable Diode Laser Absorption Spectroscopy
(TDLAS)
Measurement ranges
H
O: 0 to 20, 0 to 50, 0 to 100 lbs/MMscf
0 to 422, 0 to 1055, 0 to 2110 ppmv
CO
: 0 to 5%, 0 to 10%, 0 to 20%
Repeatability
H
O: ±1 ppmv or ±1% of reading (whichever is
greater)
CO
: ±400 ppmv or ±2% of reading (whichever
is greater)
Accuracy
H
O: ±2 ppmv or ±2% of reading (whichever is
greater)
Application data
Ambient
temperature range
-20 °C to 50 °C (-4 °F to 122 °F)
-10 °C to 60 °C (14 °F to 140 °F) —
optional