XA02744C
TDLAS Gas Analyzer
26
Hauser
If the analyzer is to be lifted by hand, designate multiple individuals to lift by
the mounting brackets, and distribute the weight among personnel to avoid
injury.
Mounting the Analyzer
The TDLAS Gas Analyzer analyzer is manufactured for wall or Unistrut
(or
equivalent) metal framing installations.
Opening/Closing the Analyzer Enclosure
Figure 4–1 SS500XP and SS2000XP lifting points
Lifting points
Lifting points
When mounting the analyzer, be sure to position the
instrument so that it is not difficult to operate adjacent
devices.
It is critical to mount the analyzer so that the inlet and outlet
lines reach the inlet and outlet connections on the chassis
while still maintaining flexibility so that the sample lines are
not under excessive stress.
Hauser analyzers are designed for operation within
the specified ambient temperature range of -20 °C to 50 °C
(-4 °F to 122 °F). Intense sun exposure in some areas may
cause the analyzer temperature to exceed the maximum.
Hazardous voltage and risk of electric shock. Failure to
properly ground the analyzer may create a high-voltage
shock hazard.