Troubleshooting
RIA15
22
Hauser
8.2
Diagnostic messages
Diagnostic
number
Short text
Corrective measure
Status signal
Diagnostic behavior
Diagnostics for the sensor
F100
Sensor error
• Check electrical wiring
• Check sensor
• Check sensor settings
F
Alarm
S901
Input signal too
small
• Check transmitter output
for defect and conformity
error
• Check transmitter for
incorrect configuration
M
Warning
S902
Input signal too
large
M
Warning
Diagnostics for the electronics
F261
Electronics
module
Replace electronics
F
Alarm
F283
Memory content
• Restart device
• Reset device
• Replace electronics
F
Alarm
F431
Factory
calibration
Replace electronics
F
Alarm
Diagnostics for the configuration
M561
Display overshoot Check scaling
M
Warning
Содержание RIA15