Troubleshooting
Batch Controller RA33
46
Hauser
9.2
Error messages
Error
Description
Remedial action
F041
Open circuit
Input current
≤
2 mA
• Incorrect wiring
• Full scale value of the measuring range configured incorrectly
• Sensor defective
• Check wiring
• Enlarge measuring range (change scaling)
• Replace sensor
M102
Over range
Input current > 3.6 mA to
≤
3.8 mA
Enlarge measuring range (change scaling)
M103
Under range
Input current
≥
20.5 mA to < 21 mA
Enlarge measuring range (change scaling)
F104
Sensor error
Input current > 2 to
≤
3.6 mA or
≥
21mA (or 22mA at 0 to 20
mA signal)
• Incorrect wiring
• Full scale value of the measuring range configured incorrectly
• Sensor defective
Pulse input >12.5 kHz or > 25 Hz
• Check wiring
• Enlarge measuring range (change scaling)
• Replace sensor
• Select a larger value for pulse value
F201
Device error (operating system error)
Contact service
F261
System error (miscellaneous hardware errors)
Contact service
F301
Setup defective
Reconfigure the device. If the error recurs, contact service
M302
Setup has been loaded from backup.
No effect on operation. To be safe, check setup (configuration)
and adjust if necessary
F303
Device data defective
Contact service
M304
Device data defective. The system continues working with
backup data.
No action required.
F305
Counters defective
Counter value is reset automatically to 0
M306
Counter defective, but system could continue working with
backup.
Check plausibility of the counter reading (compare to last stored
counter reading)
F307
Customer preset values defective
Save configuration parameters.
F309
Invalid date/time (e.g. GoldCap was empty)
Device was switched off too long. Date/time must be reconfig-
ured.
F310
Setup could not be stored
Contact service
F311
Device data could not be stored
Contact service
F312
Calibration data could not be stored
Contact service
M313
FRAM has been defragmented
No action required.
F314
Activation code is no longer correct (serial number/program
name is incorrect).
Enter new code
M315
No IP address could be obtained from the DHCP server!
Check network cable, contact network administrator.
M316
No MAC-Address or incorrect address
Contact service
F431
Calibration data missing
Contact service