Prothermo NMT 538
Hauser Systems & Gauging
27
5.1.6
Setting parameter on the matrix
The connected temp. element data must be input as an initial parameter setting to operate
NMT538 properly.
Note!
The access code 51 will be required for further parameter settings
QOG 3 Initial parameter setting of NMT538
Note!
The choice of "EQUAL" or "UNEQUAL" element interval will determine the rest of following
parameter settings.
SERVICE
TOTAL NO ELEMENT
TOTAL NO ELEMENT
KIND OF INTERVAL
16
10
EQUAL
SET DATA NMT
TEMPERATURE
MORE FUNCTION
MORE FUNCTION
MATRIX OF
MATRIX OF
>GROUP SELECT<: E
>GROUP SELECT<: E
>GROUP SELECT<: E
Enter "E" key several times
until "MORE FUNCTION" appears
again
Enter "+" key 5 times to display
"SET DATA NMT""
Enter "+" key 3 times to display
"TOTAL NO ELEMENT"
Enter desired element number
with "+" or "-" keys (ie,10 in this case)
then, "E" to confirm
Enter "E" key twice, go to
"KIND OF INTERVAL"
Use "+" or "-" key to select
"EQUAL" or "UNEQUAL", then
"E" to confirm.
Enter "E" key several times to exit
"SET DATA NMT" row.
"E" key once to enter the selection
of Dynamic matrix (G1 to G9)
Enter with "+" or "-" key, going to
"TEMPERATURE" matrix
Caution!!
NMT538 must be configured at
the NMS53x matrix GVH 362
"CONNECTION NMT" prior to
perform parameter setting
Now you are in the
G4 "TEMPERATURE" matrix
Note!
Choice of "UNEQUAL" allows specific
element position setting
"EQUAL" automatically establish
each element position by setting
"BOTTOM POINT" &
"ELEMENT INTERVAL"
Default setting would be 16 (max.
available number)
LCD display
Procedure
Remark
5 Operation with the Proservo NMS53x