background image

BA099D/06/en/03.08
71070185

Operating Instructions

Promass

Heating jacket for Promass A, E, H, I, M, S, P

Содержание Promass

Страница 1: ...BA099D 06 en 03 08 71070185 Operating Instructions Promass Heating jacket for Promass A E H I M S P ...

Страница 2: ......

Страница 3: ...stalling the heating jacket 8 3 3 1 Mounting kit 8 3 3 2 Mount the isolation 9 3 3 3 Connecting to in house heating pipes 9 3 4 Post connection check 9 4 Technical data 10 4 1 Technical data at a glance 10 4 1 1 Application 10 4 1 2 Function and system design 10 4 1 3 Operating conditions 10 4 1 4 Operating conditions environment 10 4 1 5 Operating conditions Process 10 4 1 6 Mechanical constructi...

Страница 4: ...the remote version at very high fluid and heating temperatures we recommend the use of the long necked version remote version for heating so that the ambient temperature of the connection housing and the connecting cable is not exceeded With the compact version make sure you take appropriate measures orientation etc so that the maximum ambient temperature for the transmitter is not exceeded otherw...

Страница 5: ...ted crevices or diffused through plastic Costs incurred for waste disposal and injury burns etc due to inadequate cleaning will be charged to the owner operator 1 5 Notes on safety conventions and icons The heating jackets are designed to meet state of the art safety requirements have been tested and left the factory in a condition in which they are safe to operate They can however be a source of ...

Страница 6: ...ssure heating medium 6 Maximum permitted temperature heating medium 7 Corresponding sensor 2 2 Certificates and approvals The devices are designed to meet state of the art safety requirements have been tested and left the factory in a condition in which they are safe to operate The devices correspond to article 3 3 of the EEC guideline 97 23 BG Pressure Equipment Directive and are laid out and man...

Страница 7: ...g electronics are not exceeded Page 10 With additional heating jacket insulation always keep to the minimum safety distances to the device Page 13 ff Keep to the following ambient temperatures depending on the design of the measuring system The material load curves pressure temperature diagrams for the heating jacket are to be found on Page 12 Device version Max ambient temperatures Compact versio...

Страница 8: ...P Join the two plates of the heating jacket by nuts and bolts Warning With large nominal diameters a support may be required due to the heavy weight of the sensor and the heating jacket Never use the feed and return pipes of the heating system to support the heating jacket The weight data for the sensors are listed in the Technical data of the corresponding Operating Instructions Vertical Horizont...

Страница 9: ...essure is maintained even if the total flow is reduced but is passed on almost completely to existing still open valves or pipe restrictions due to the greatly reduced pipe loss 3 4 Post connection check Perform the following checks after completing the installation Device condition and specifications Notes Is the device damaged visual inspection Does the device correspond to specifications at the...

Страница 10: ...ating jacket consists of a heating plate Promass H E S P The heating jacket consists of two heating plate 4 1 3 Operating conditions Installation instructions Installation position arbitrary horizontal or vertically restrictions and further references Page 7 4 1 4 Operating conditions environment Ambient temperature Transmitter or connection housing All versions max 60 C 140 F Detailed description...

Страница 11: ...g h A H S P E I M 1 400 2 450 4 500 8 1600 400 700 350 15 1560 440 700 350 25 1520 500 650 350 40 1410 610 560 670 50 1200 850 450 1000 80 1500 differencial pressure 1bar DN Heat transfer coefficient α in C h A H S P E I M 1 150 2 120 4 80 8 210 250 65 220 15 205 235 65 220 25 200 220 60 200 40 180 180 50 165 50 160 100 40 110 80 25 uninsulated ...

Страница 12: ... 50 to 350 C 58 to 662 F But the maximal permissible fluid temperature always depends on sensor A0006939 ae Fig 2 Material load through the heating medium pressure temperature diagram 4 1 7 Certificates and approvals Pressure device approval The devices correspond to article 3 3 of the EEC guideline 97 23 BG Pressure Equipment Directive and are laid out and manufactured after good engineer practic...

Страница 13: ...41 Abb 3 Heating jacket Promass A R Insulation customer supply e g mineral wool S Safety distance min 20 mm T Process connection of the heating jacket acc to tabele E s c E A B C S D R R T G E F DN A B C D E F G Weight kg Volume liter 1 245 176 34 15 210 18 93 5 1 5 0 1 2 308 176 34 15 265 18 93 5 2 0 0 1 4 435 210 34 15 391 20 112 5 3 0 0 2 Dimensions in mm ...

Страница 14: ...sc E S R T E B C D R T F A DN A B C D E F Weight1 kg Volume1 liter 8 146 150 62 130 100 39 0 6 0 05 15 189 160 62 130 140 42 5 1 0 05 25 240 170 62 130 190 43 1 3 0 1 40 336 190 76 144 282 45 2 0 15 50 430 255 96 168 368 5 52 3 5 0 2 Dimensions in mm 1 per Jacket DN A B C D E F Weight1 kg Volume1 liter 8 189 160 92 160 140 42 5 1 0 05 15 320 170 92 160 268 43 1 7 0 1 25 450 190 92 160 396 45 2 7 0...

Страница 15: ...al wool S Safety distance min 20 mm T Process connection of the heating jacket acc to tabele Promass M Esc E C A B T R E D S T G 2 F 1 H F F DN A B C D E F G H Weight1 kg Volume1 liter 8 255 121 5 79 130 40 87 5 200 1 0 05 15 285 134 5 82 130 40 102 5 230 40 1 0 05 25 310 141 5 89 134 40 115 255 50 1 2 0 05 40 410 156 5 107 5 138 50 155 354 80 2 0 1 50 542 169 5 125 152 50 221 480 115 3 0 2 80 642...

Страница 16: ... 0 1 15 290 170 5 125 132 40 1 7 230 30 1 5 0 1 15 FB 428 176 5 125 132 50 2 5 370 74 2 5 0 1 25 428 176 5 125 132 50 2 5 370 74 2 5 0 1 25 FB 530 202 151 136 50 3 8 470 120 3 7 0 15 40 530 202 151 136 50 3 8 470 120 3 7 0 3 40 FB 645 232 180 132 50 5 5 580 182 5 5 3 0 4 50 645 232 180 132 50 5 5 580 182 5 5 3 0 4 Dimensions in mm FB full bore version 1 per Jacket ...

Страница 17: ...11 9 G 1 5 16 AF SW 33 51 3 19 1 15 7 1 Dimensions in mm 2 DIN EN ISO 228 S ISO D L E F G Flange according to ASME B16 5 1 4404 316L 316 Pressure stage ISO2 D1 E1 F1 G1 L1 S Cl 150 G 15 7 60 5 88 9 4 Ø15 7 87 8 SW 21 7 8 AF G 15 7 60 5 88 9 4 Ø15 7 87 8 Cl 300 G 15 7 66 5 95 2 4 Ø15 7 92 3 G 15 7 66 5 95 2 4 Ø15 7 92 3 Flange EN 1092 1 DIN 2501 PN 40 1 4404 316L 316 Pressure stage ISO2 D1 E1 F1 G1...

Страница 18: ...t Promass A R Insulation customer supply e g mineral wool S Safety distance min 0 79 T Process connection of the heating jacket acc to tabele E s c E A B C S D R R T G E F DN A B C D E F G Weight lbs Volume US gal 1 24 9 65 6 93 1 34 0 59 8 27 0 71 3 68 3 3 0 03 1 12 12 13 6 93 1 34 0 59 10 43 0 71 3 68 4 4 0 03 1 8 17 13 8 27 1 34 0 59 15 39 0 79 4 43 6 6 0 05 Dimensions in inch ...

Страница 19: ...1 US gal 3 8 5 75 5 91 2 44 5 12 3 94 1 54 1 32 0 01 1 2 7 44 6 30 2 44 5 12 5 51 1 67 2 21 0 01 1 9 45 6 69 2 44 5 12 7 48 1 69 2 87 0 03 1 13 23 7 48 2 99 5 67 11 10 1 77 4 41 0 04 2 16 93 10 04 3 78 6 61 14 51 2 05 7 72 0 05 Dimensions in inch 1 per Jacket DN A B C D E F Weight 1 lbs Volume 1 US gal 3 8 7 44 6 30 3 62 6 30 5 51 1 67 2 21 0 01 1 2 12 60 6 69 3 62 6 30 10 55 1 69 3 75 0 03 1 17 7...

Страница 20: ...rocess connection of the heating jacket acc to tabele Promass M Esc E C A B T R E D S T G 2 F 1 H F F DN A B C D E F G H Weight 1 lbs Volume 1 US gal 3 8 10 04 4 78 3 11 5 12 1 57 3 44 7 87 2 21 0 13 1 2 11 22 5 30 3 23 5 12 1 57 4 04 9 06 1 57 2 21 0 13 1 12 20 5 57 3 50 5 28 1 57 4 53 10 04 1 97 2 65 0 13 1 16 14 6 16 4 23 5 43 1 97 6 10 13 94 3 15 4 41 0 03 2 21 34 6 67 4 92 5 98 1 97 8 70 18 9...

Страница 21: ...7 0 07 9 06 1 18 3 31 0 03 1 2 FB 16 85 6 95 4 92 5 20 1 97 0 10 14 57 2 91 5 51 0 03 1 16 85 6 95 4 92 5 20 1 97 0 10 14 57 2 91 5 51 0 03 1 FB 20 87 7 95 5 94 5 35 1 97 0 15 18 50 4 72 8 16 0 04 1 20 87 7 95 5 94 5 35 1 97 0 15 18 50 4 72 8 16 0 08 1 FB 25 39 9 13 7 09 5 20 1 97 0 22 22 83 7 19 11 69 0 11 2 25 39 9 13 7 09 5 20 1 97 0 22 22 83 7 19 11 69 0 11 Dimensions in inch FB full bore vers...

Страница 22: ...5 0 47 G 1 5 16 AF 2 02 0 75 0 62 1 Dimensions in inch 2 DIN EN ISO 228 S ISO D L E F G Flange according to ASME B16 5 1 4404 316L 316 Pressure stage ISO2 D1 E1 F1 G1 L1 S Cl 150 G 0 62 2 38 3 5 4 Ø0 62 3 46 7 8 AF G 0 62 2 38 3 5 4 Ø0 62 3 46 Cl 300 G 0 62 2 62 3 75 4 Ø0 62 3 63 G 0 62 2 62 3 75 4 Ø0 62 3 63 Flange EN 1092 1 DIN 2501 PN 40 1 4404 316L 316 Pressure stage ISO2 D1 E1 F1 G1 L1 S PN 4...

Страница 23: ...____________________________________________ ______________________________________________________________________________________________________________ We hereby certify that this declaration is filled out truthfully and completely to the best of our knowledge We further certify that the returned parts have been carefully cleaned To the best of our knowledge they are free of any residues in da...

Страница 24: ...www endress com worldwide BA099D 06 en 03 08 71070185 FM SGML6 0 ProMoDo ...

Отзывы: