TI069D/24/ae

Technical Information

Proline t-mass 65F, 65I

Thermal mass flow measuring system
Direct mass flow measurement of gases

Application

For measuring the mass flowrate of a wide range of gas 
types e.g.

• Compressed air
• Natural gas flow to boilers / dryers
• Carbon Dioxide flow in breweries
• Biogas and aeration air in waste water plants
• Gas production (e. g. Ar, N

2

, CO

2

, He, O

2

)

• Leakage detection

Approvals for hazardous area:
• ATEX, FM, CSA

Connection to all common process control systems:
• HART

®

, Profibus

®

 DP, MODBUS

®

 RS485

Your benefits

Direct measurement of gas mass flow. Provides tempera-
ture as an output.

The 

Proline transmitter concept

 comprises:

• Modular device and operating concept resulting in a 

higher degree of efficiency

• Quick setup operating menus for ease of commission-

ing

• On board software freely allows the selection of up to 

20 pure gases and creation of mixed gases with a
maximum of 8 components (e. g. Biogas)

The 

t-mass sensors

 offer:

• Negligible pressure drop or loss
• Wide turndown of up to 100:1
• Insertion version can be programmed for circular pipe 

or rectangular ducting installation

• Each device individually calibrated and delivered with 

a traceable certificate

• Can be calibrated with flow conditioner on request.
• Optional cold tap device for insertion allowing ease of 

removal/replacement for low pressure and non-toxic 
gas applications.

Содержание Proline t-mass 65F

Страница 1: ...empera ture as an output The Proline transmitter concept comprises Modular device and operating concept resulting in a higher degree of efficiency Quick setup operating menus for ease of commission ing On board software freely allows the selection of up to 20 pure gases and creation of mixed gases with a maximum of 8 components e g Biogas The t mass sensors offer Negligible pressure drop or loss W...

Страница 2: ...et runs 12 Insertion version 16 Length of connecting cable 18 Operating conditions Environment 18 Ambient temperature range 18 Storage temperature 18 Degree of protection 18 Shock resistance 18 Vibration resistance 18 Electromagnetic compatibility EMC 18 Operating conditions Process 19 Medium temperature range 19 Pressure loss 19 Medium pressure range nominal pressure 19 Medium pressure 19 Mechani...

Страница 3: ... required to maintain the differential temperature The measured heater power is therefore a measure of the gas mass flowrate a0005136 Measuring system The measuring system consists of a transmitter and a sensor Two versions are available Compact version transmitter and sensor form a single mechanical unit Remote version transmitter and sensor are mounted physically separate from one another Transm...

Страница 4: ...mited to a value below 70 m sec DN kg h Nm3 h at 0 C 1 013 bar a scf min at 15 C 1 013 bar a min max min max min max 15 0 5 53 0 38 41 0 23 25 25 2 200 1 5 155 1 0 96 40 6 555 4 6 429 3 0 266 50 10 910 7 7 704 5 0 436 80 20 2030 15 5 1570 10 974 100 38 3750 29 2900 18 1800 DN lb h Sm3 h at 59 F 14 7 psi a scf min at 59 F 14 7 psi a minimum maximum minimum maximum minimum maximum 1 2 1 1 116 0 4 42...

Страница 5: ...nits DN lb h Sm3 h at 59 F 14 7 psi a scf min at 59 F 14 7 psi a minimum maximum minimum maximum minimum maximum 3 44 4466 16 1657 9 926 4 84 8250 30 3060 17 1710 6 110 16500 40 6120 22 8 3420 8 176 27500 64 10200 36 5 5700 10 264 44000 98 16300 55 9120 12 396 61000 146 22855 82 12768 16 660 110000 245 40810 136 22800 20 1100 176000 408 65300 228 36480 24 1540 253000 570 93870 319 52440 28 1980 34...

Страница 6: ...Passive open collector 30 V DC 250 mA galvanically isolated Frequency output full scale frequency 2 to 1000 Hz fmax 1250 Hz on off ratio 1 1 pulse width max 2s time constant selectable 0 0 to 100 0 s Pulse output pulse value and pulse polarity can be selected pulse width adjustable 0 5 to 2000 ms PROFIBUS DP interface PROFIBUS DP in accordance with IEC 61158 galvanically isolated Profile version 3...

Страница 7: ... Status and alarm messages in accordance with PROFIBUS Profile Version 3 0 MODBUS RS485 If an error occurs the value NaN not a number is output for the measured values Load see Output signal Low flow cut off Switch points for low flow cut off are programmable Galvanic isolation All circuits for inputs outputs and power supply are galvanically isolated from each other Switching output Relay output ...

Страница 8: ... PROFIBUS DP MODBUS RS485 B RxD TxD N e Ground terminal for signal cable shield RS485 line Observe that the stripped and twisted lengths of cable shield to the ground terminal are as short as possible max 5 mm f Service connector for connecting service interface FXA193 Fieldcheck Tof Tool Fieldtool g Signal cable See Terminal assignment Page 7 PROFIBUS cable for external termination optional Termi...

Страница 9: ...on boards permanent assignment 65 A Frequency output Current output HART 65 B Relay output Relay output Frequency output Current output HART 65 J 5V ext termination PROFIBUS DP 65 Q Status input MODBUS RS485 65 R Current output 2 Ex i active Current output 1 Ex i active HART 65 S Frequency output Ex i passive Current output Ex i Active HART 65 T Frequency output Ex i passive Current output Ex i Pa...

Страница 10: ... 260 V 18 2 W 20 to 55 V 14 W including sensor DC 8 W including sensor Power supply failure Lasting minimum 1 power cycle EEPROM HistoROM T DAT saves measuring system data if the power supply fails HistoROM S DAT is on exchangeable data storage chip with sensor specific data pipe type nominal diam eter serial number flow conditioner zero point etc Totalizer stops at the last value determined Poten...

Страница 11: ...nd the sensing elements e g do not install the sensor at a low point in the installation without adequate drainage Make sure that the direction arrow on the sensor matches the direction of flow direction of fluid flow through the pipe Caution When mounting an insertion meter in a vertical line it is recommended that a remote version is used or support is provided for the compact version Flanged ve...

Страница 12: ...n When disturbances e g pipe elbows reducers valves T pieces etc are located upstream of the thermal meter precautions must be taken to minimise any effects on the measuring performance The figure on the following page illustrates the minimum recommended inlet and outlet runs expressed in mul tiples of the pipe diameter longer lengths should always be used if they are available in the metering run...

Страница 13: ...e where possible a modulating control valve should be mounted downstream of a flowmeter a Inlet run b Outlet run Caution Wherever possible mount any control and butterfly shut off valves downstream of flowmeter 15 x Dia Dia Dia Dia Dia Dia Dia Dia Dia Dia Dia Dia Dia 2 x Dia 2 x Dia 2 x Dia 2 x Dia 2 x Dia 2 x Dia 5 x Dia 5 x Dia 5 x Dia 5 x Dia 5 x Dia 5 x a 1 b 3 a b 15 x 5 a b 20 x 4 a b 35 x 6...

Страница 14: ...ge of pipe diameters it is not possible to calibrate insertion meters with a flow conditioner if ordered together The conditioners used are based on the Mitsubishi design which are available from line sizes 80 mm 3 to 300 mm 12 Flow conditoner for use with flanged sensors This is a special Endress Hauser version designed specially for use with the t mass F sensor Sizes DN 25 to 100 1 to 4 The flow...

Страница 15: ...ssure loss calculation a0005243 en Pressure loss Δp in mbar or psi Mass flow g in kg h or lbm h Density ρ in kg m3 or lbm ft3 Diameter D in mm or inch Constant 1876 metric or 8 4 10 7 US units Metric units calculation g 148 kg h ρ 5 94 kg m3 at 5 bar abs at 20 C D 28 5 mm for a DN 25 PN 40 meter a0005244 en US units calculation g 326 lbm h ρ 0 37 lbm ft3 at 72 5 psia at 68 F D 1 05 inch for a 1 Cl...

Страница 16: ...ss C Height of the mounting boss on the pipe or duct including the sensor fitting and cold hot tap if fitted The tube section is supplied with a graduated scale along its length calibrated in millimeters or inches It is important that the sensor be installed so that the top of the adjustable fitting is aligned with the value on the scale that is equal to the following calculated value For pipe dia...

Страница 17: ...unted at 90 degrees to the flow direction Any deviation from this angle in any plane may cause flow disturbances around the measuring point that could cause errors Flow direction alignment It is very important that the sensor is aligned correctly with the direction of flow There are two guidelines for correct alignment The arrows on the sides of the sensor housing assembly are pointing in the same...

Страница 18: ... F available on request Note Install the device in a shaded location Avoid direct sunlight particularly in warm climatic regions a protective hood is available on request At ambient temperatures below 20 C 4 F the readability of the display may be impaired It must be remembered that the sensor operating principle is based on a heat loss mechanism therefore the sensor performs best when the ambient...

Страница 19: ...e loss 2 mbar 0 029 psi maximum without flow conditioner Medium pressure range nom inal pressure t mass F 0 5 to 40 bar gauge 7 25 to 580 psig t mass I 0 5 to 20 bar gauge 7 25 to 290 psig Medium pressure Pressure temperature curve to EN DIN stainless steel a0005240 Pressure temperature curve to JIS B2238 and ANSI 16 5 stainless steel a0005241 200 100 0 5 10 15 20 25 30 35 40 45 0 100 200 300 400 ...

Страница 20: ...nd the permissible range 20 to 60 C 4 to 140 F optional 40 to 60 C 40 to 140 F Install the device in a shaded location Avoid direct sunlight Always install the wall mount housing in such a way that the cable entries are pointing down A B C D E F G H J K L M N O P Q R 215 250 90 5 159 5 135 90 45 50 81 53 95 53 102 81 5 11 5 192 8xM5 A B C D E F G H J K L M N O P Q R 8 46 9 84 3 56 6 28 5 32 3 54 1...

Страница 21: ...e mounting and installation in control panel a0005256 ae A Pipe mounting of wallmount housing B Installation in control panel of wallmount housing Caution If a warm pipe is used for installation make sure that the housing temperature does not exceed the maximum permitted value of 60 C 140 F a b c c 90 3 54 35 1 38 192 7 56 81 5 3 2 mm inch A B 245 9 65 0 5 0 5 210 8 27 Ø 20 to 70 0 78 to 2 75 110 ...

Страница 22: ... ir c u i t s a r e a l i v e Nicht eigensichere Stromkreise durch IP40 Abdeckung geschützt Non intrinsically safe circuits Ip40 protected Boucles de courant sans sécurité intrinsèque protégées par Ip40 347 13 7 180 7 1 167 6 57 100 3 94 144 5 67 100 3 94 130 5 12 217 8 54 240 9 44 242 9 52 265 10 4 Ø 8 6 M8 0 39 186 7 32 mm inch Nicht unter Spannung öffnen K e e p c o v e r t i g h t w h i l e c ...

Страница 23: ...sions Remote sensor housing flanged version II2GD zone 1 a0005158 ae For dimension V of the flanged remote sensor version please refer for the metric units to Page 24 and for the US units to Page 28 102 4 0 Ø 129 5 1 52 2 0 79 3 1 50 1 96 V mm inch ...

Страница 24: ... DIN 2501 DIN 2512N 1 PN 40 1 4404 316L 316 Surface roughness flange EN 1092 1 Form B1 DIN 2526 Form C Ra 6 3 to 12 5 μm DN di B G L LK N P S T U V 15 13 9 276 5 95 245 65 4 x Ø14 132 5 16 112 5 17 3 116 5 25 24 3 276 5 115 245 85 4 x Ø14 132 5 18 112 5 28 5 116 5 40 38 1 173 5 150 320 110 4 x Ø18 200 18 120 43 1 113 5 50 49 2 278 5 165 400 125 4 x Ø18 250 20 150 54 5 118 5 80 73 7 291 200 640 160...

Страница 25: ...04 316L 316 Surface roughness flange Ra 3 2 to 6 3 μm DN di B G L LK N P S T U V 15 13 9 276 5 95 245 70 4 x Ø15 132 5 14 112 5 16 1 116 5 25 24 3 276 5 125 245 90 4 x Ø19 132 5 16 112 5 27 2 116 5 40 38 1 273 5 140 320 105 4 x Ø19 200 18 120 41 2 113 5 50 49 2 278 5 155 400 120 8 x Ø19 250 18 150 52 7 118 5 80 73 7 291 0 200 640 160 8 x Ø23 400 22 240 78 1 131 0 100 97 303 0 225 800 185 8 x Ø23 5...

Страница 26: ...s Remote sensor housing insertion version a0005159 ae Insertion version length V W L LR 235 235 244 5 467 5 409 5 335 335 344 5 567 5 509 5 435 435 444 5 667 5 609 5 608 608 617 5 840 5 782 5 G 1 A 1 MNPT L W V 227 8 94 207 8 15 168 6 61 mm inch 160 6 3 55 2 2 63 2 5 G 1 A 1 MNPT LR W V Ø 129 5 1 52 2 0 79 3 1 50 1 96 102 4 0 55 2 2 63 2 5 mm inch ...

Страница 27: ... 17 5 2 5 7 0 0 7 50 A 123 150 17 5 2 5 8 5 1 0 80 C 158 185 17 5 2 5 13 0 2 3 100 C 187 220 22 5 2 5 17 0 4 1 17 0 4 1 JIS 10K 20K Sched 40 Sched 80 DN Type D1 D2 F Q R S Weight kg S Weight kg 25 A 87 115 17 5 2 5 4 5 0 4 4 0 0 4 40 A 102 130 17 5 2 5 6 5 0 7 6 0 0 7 50 B 117 145 17 5 2 5 8 5 1 2 8 0 1 1 80 C 157 188 21 5 2 5 12 5 3 0 12 0 2 8 100 C 182 214 21 5 2 5 16 5 5 1 15 5 4 8 D1 D2 Ø F S ...

Страница 28: ...11 46 7 50 25 2 6 00 4 x Ø 3 4 15 7 0 94 9 45 3 07 5 16 4 4 29 11 93 9 00 31 5 7 50 8 x Ø 3 4 19 7 0 94 11 8 4 03 5 63 Flange ANSI B16 5 Cl 150 Sched 80 1 4404 316L 316 Surface roughness flange N9 250 μinch Dia di B G L LK N P S T U V 1 2 0 87 10 88 3 50 9 6 2 38 4 x Ø 5 8 5 22 0 44 4 43 0 55 4 59 1 1 28 10 88 4 25 9 6 3 12 4 x Ø 5 8 5 22 0 56 4 43 0 96 4 59 1 1 2 1 74 10 77 5 00 12 6 3 88 4 x Ø 5...

Страница 29: ...9 11 46 8 25 25 2 6 62 8 x Ø 7 8 15 7 1 12 9 45 3 07 5 16 4 4 29 11 93 10 00 31 5 7 88 8 x Ø 7 8 19 7 1 25 11 8 4 03 5 63 Flange ANSI B16 5 Cl 300 Sched 80 1 4404 316L 316 Surface roughness flange N9 250 μinch Dia di B G L LK N P S T U V 1 2 0 87 10 88 3 75 9 6 2 62 4 x Ø 5 8 5 22 0 56 4 43 0 55 4 59 1 1 28 10 88 4 88 9 6 3 50 4 x Ø 3 4 5 22 0 69 4 43 0 96 4 59 1 1 2 1 74 10 77 6 12 12 6 4 50 4 x ...

Страница 30: ...Remote sensor housing insertion version a0005159 ae Insertion length version V W L LR 9 25 9 25 9 63 18 4 16 1 13 2 13 2 13 6 22 3 20 1 17 1 17 1 17 5 26 3 24 0 23 9 23 9 24 3 33 1 30 8 G 1 A 1 MNPT L W V 227 8 94 207 8 15 168 6 61 mm inch 160 6 3 55 2 2 63 2 5 G 1 A 1 MNPT LR W V Ø 129 5 1 52 2 0 79 3 1 50 1 96 102 4 0 55 2 2 63 2 5 mm inch ...

Страница 31: ... Sched 80 Dia Type D1 D2 D3 F Q R S Weight lb S Weight lb 1 A 3 36 4 33 0 67 0 2 0 1 0 18 0 8 0 16 0 7 1 1 2 A 4 31 5 51 0 85 0 2 0 1 0 26 1 9 0 24 1 7 2 D 4 80 5 90 4 55 0 75 0 2 0 1 0 33 2 8 0 31 2 7 3 D 6 41 7 67 5 67 0 87 0 2 0 1 0 49 7 0 0 42 6 8 4 C 7 75 8 97 0 81 0 2 0 1 0 65 11 7 0 61 11 9 D1 D2 Ø F S A B C R Q D D2 D3 D1 ...

Страница 32: ... Sensor body DN 15 to 25 1 2 to 1 stainless steel cast CF3M A351 DN 40 to 100 1 1 2 to 4 1 4404 to EN10216 5 and 316 316L SS to A312 Flanges process connections According to EN 1092 1 DIN 2501 DIN 2512N ANSI B16 5 JIS B2238 stainless steel 1 4404 to EN 10222 5 and 316L 316 SS to A182 Transducer body 1 4404 to EN10272 and 316L SS to A479 Alloy C22 and UNS N06022 to B574 t mass F DN 15 25 40 50 80 1...

Страница 33: ...loy C22 and UNS N06022 to B626 Compression fitting 1 4404 to EN 10272 and 316 316L SS to A479 Compression fitting seal PEEK Bonded seals EPDM Kalrez Nitrile 316 316L SS outer ring Low pressure retractable insertion fitting Cold tap Lower tube section 1 4404 to EN 10272 and 316 316L SS to A479 Upper tube section 1 4404 to EN 10216 5 and 316 316L SS to A312 Ball valve 1 4408 to EN 10213 4 and CF8M S...

Страница 34: ...ces in the hazardous area Example t mass 65F PROFIBUS DP certification The flow device has successfully passed all the test procedures carried out and is certified and registered by the PNO PROFIBUS User Organization The device thus meets all the requirements of the following specifications The device can also be operated with certified devices of other manufacturers interoperability MODBUS certif...

Страница 35: ... and laboratory control equipment NAMUR NE 43 Standardisation of the signal level for the breakdown information of digital transmitters with analogue output signal NAMUR NE 53 Software of field devices and signal processing devices with digital electronics Pressure Equipment Directive Flow meters with a nominal diameter smaller or equal DN 25 1 are covered by Art 3 3 of the European directive 97 2...

Страница 36: ...or obtaining measured values via the current output HART 4 to 20 mA DXR375 Contact your local Endress Hauser sales representative for more information Accessory Description Order code Applicator Software for selecting and configuring flowmeters Applicator can be downloaded from the Internet or ordered on CD ROM for installation on a local PC DKA80 Contact your local Endress Hauser sales representa...

Страница 37: ...e Nemours Co Wilmington USA HART Registered trademark of HART Communication Foundation Austin USA PROFIBUS Registered trademark of the PROFIBUS User Organisation Karlsruhe Germany MODBUS Registered trademark of the MODBUS Organisation t mass HistoROM S DAT T DAT F CHIP ToF Tool Fieldtool Package Fieldcheck Applicator Registered or registration pending trademarks of Endress Hauser Flowtec AG Reinac...

Страница 38: ... Pressure test 4 Material certificate tube flange sensor plus pressure test 5 CRN approval only for ANSI flanges 8 CRN approval material certificate tube flange sensor plus pressure test only for ANSI flanges 070 Approval A Non hazardous N FM XP Cl I Div 1 CSA Cl I Div 1 Grps A D Zone 1 not for housing C R FM NI CL I Div 2 CSA Cl I Div 2 Grps A D Zone 2 not for housing G not for output input R S T...

Страница 39: ...G not for output input R S T U T masss 65I Insertion Version 005 010 020 030 040 050 060 070 080 090 100 110 120 T masss 65I 080 Housing A Compact aluminum NEMA 4X C Wall mount aluminum NEMA 4X G Field housing aluminum NEMA 4X not for approvals A R 090 Cable entry B 1 2 NPT X Not used sensor only only for housing A 100 Display operation 7 85 to 260 VAC without display remote configuration not for ...

Страница 40: ...D 24 ae 02 08 2008 Endress Hauser Canada Endress Hauser Canada 1075 Sutton Drive Burlington ON L7L 5Z8 Tel 905 681 9292 800 668 3199 Fax 905 681 9444 www ca endress com Mexico Endress Hauser México S A de C V Fernando Montes de Oca 21 Edificio A Piso 3 Fracc Industrial San Nicolás 54030 Tlalnepantla de Baz Estado de México México Tel 52 55 5321 2080 Fax 52 55 5321 2099 eh mexico mx endress com www...

Отзывы: