Proline t-mass 65F, 65I
Hauser
15
The flow conditioner is fitted between two piping flanges and centered with mounting bolts.
a0005116
Flow conditioner mounting arrangement
1 = Perforated plate flow conditioner
2 = Seal/gasket
3 = Notch for positioning
Note!
The flow conditioner has a notch to indicate the mounting position. Mount vertically.
Pressure loss calculation
a0005243-en
Pressure loss:
Δ
p in mbar or psi
Mass flow :
g
in kg/h or lbm/h
Density:
ρ
in kg/m
3
or lbm/ft
3
Diameter: D in mm or inch
Constant = 1876 (metric) or 8.4 · 10
-7
(US units)
Metric units calculation:
g
= 148 kg/h
ρ =
5.94 kg/m
3
(at 5 bar abs at 20 °C)
D = 28.5 mm (for a DN 25, PN 40 meter)
a0005244-en
US units calculation:
g
= 326 lbm/h
ρ =
0.37 lbm/ft
3
(at 72.5 psia at 68°F)
D = 1.05 inch (for a 1" Class 150 Sched 40 meter)
a0005244-ae
1
2
2
3
D
p = constant
g
2
r
D
4
1
D
p = 1876
1
148
2
5.94
28.5
4
= 10.5 mbar
D
p = 8.4
1
326
2
0.371
1.05
4
= 0.198 psi
10
-7