Commissioning
Proline t-mass 65 PROFIBUS DP/PA
64
Hauser
6.6
PROFIBUS DP cyclic data transmission
Below is a description of the cyclic data transmission when using the t-mass GSD file (complete
device functionality).
6.6.1
Block model
The block model illustrated shows which input and output data the measuring device provides for
cyclic data transmission via PROFIBUS DP.
a0005482-en
Fig. 38:
Block model for t-mass PROFIBUS DP Profile 3.0
Esc
E
+
-
XXX.XXX.XX
Measured variable 1
Measured variable 2
Measured variable 3
Measured variable …
Measured variable 1
Measured variable 2
Measured variable 3
Measured variable …
Analog Input 1
Output value AI
Output value AI
Analog Input 2
Signal
processing
PROFIBUS
PROFILE
Parameter
Manufacturer
specific
Parameter
Control (CONTROL_BLOCK)
Display value (DISPLAY_VALUE)
Transducer
block
Totalizer 2
Out value TOTAL
Local
Display
Out value TOTAL
Totalizer 1
Process pressure (PRESSURE_VALUE)
PROFIBUS DP
Physical
Block
Configuration SETTOT, MODETOT
Configuration SETTOT, MODETOT
Analog Input 3
Output value AI