Commissioning
Proline Promass E 200
44
Hauser
Wizard
Meaning
Select medium
Defining the medium
Current output 1
Configuring output 1
Current output 2
Configuring output 2
Pulse/frequency/switch output
Configuring the selected output type
Display
Configuring the measured value display
Output conditioning
Defining the output conditioning
Low flow cut off
Configuring the low flow cut off
Partial filled pipe detection
Configuring the monitoring of partial and empty pipe
detection
10.5
Defining the tag name
To enable fast identification of the measuring point within the system, you can enter a unique
designation using the
Device tag
parameter and thus change the factory setting.
Navigation
"Setup" menu → Device tag
Parameter overview with brief description
Parameter
Description
User entry
Factory setting
Device tag
Enter tag for measuring
point.
Max. 32 characters, such as
letters, numbers or special
characters (e.g. @, %, /).
Promass
10.6
Protecting settings from unauthorized access
The following options exist for protecting the configuration of the measuring device from
unintentional modification after commissioning:
• Write protection via access code (→ 44)
• Write protection via write protection switch (→ 45)
• Write protection via keypad lock (→ 39)
10.6.1
Write protection via access code
With the customer-specific access code, the parameters for the measuring device
configuration are write-protected and their values can no longer be changed via local
operation.
Navigation
"Setup" menu → Advanced setup → Administration → Def. access code
Structure of the submenu
Define access code
→
Define access code
Confirm access code