Proline Promass E 200
Electrical connection
Hauser
27
An external power supply is required for each output. The following supply voltage values
apply for PROFIBUS PA and the pulse/frequency/switch output:
Order code for "Output"
Minimum terminal
voltage
Maximum terminal
voltage
Option
A
: 4-20 mA HART
For 4 mA: ≥ DC 18 V
For 20 mA: ≥ DC 14 V
DC 35 V
Option
B
: 4-20 mA HART, pulse/frequency/switch
output
For 4 mA: ≥ DC 18 V
For 20 mA: ≥ DC 14 V
DC 35 V
Option
C
For 4 mA: ≥ DC 18 V
For 20 mA: ≥ DC 14 V
DC 30 V
Option
G
: PROFIBUS PA, pulse/frequency/switch output 9 V
DC 32 V
1)
External supply voltage of the power supply unit with load.
2)
For device versions with SD03 local display: The terminal voltage must be increased by DC 2 V if backlighting is
used.
3)
For device version with SD03 local display: The terminal voltage must be increased by DC 0.5 V if backlighting is
used.
Load
Load for current output: 0 to 500 Ω, depending on the external supply voltage of the power
supply unit
Calculation of the maximum load
Depending on the supply voltage of the power supply unit (U
S
), the maximum load (R
B
)
including line resistance must be observed to ensure adequate terminal voltage at the device.
In doing so, observe the minimum terminal voltage (→ 26)
• For U
S
= 18 to 18.9 V: R
B
≤ (U
S
- 18 V) :0.0036 A
• For U
S
= 18.9 to 24.5 V: R
B
≤ (U
S
- 13.5 V) :0.022 A
• For U
S
= 24.5 to 30 V: R
B
≤500 Ω