Brief Operating Instructions

Proline Promass E 200

Coriolis flowmeter

These Instructions are Brief Operating Instructions; they are

not a substitute for the Operating Instructions pertaining to

the device.
Detailed information about the device can be found in the

Operating Instructions and the other documentation:

• On the CD-ROM supplied (is not included in the delivery for

all device versions).

• Available for all device versions via:

– Internet: 

www.endress.com/deviceviewer

– Smart phone/tablet: 

Hauser Operations App

Description of the procedure (→   10)

KA00050D/06/EN/16.14

71241802

Содержание Proline Promass E 200

Страница 1: ...ons pertaining to the device Detailed information about the device can be found in the Operating Instructions and the other documentation On the CD ROM supplied is not included in the delivery for all...

Страница 2: ...conditions 14 6 2 Mounting the measuring device 19 6 3 Post installation check 21 7 Electrical connection 22 7 1 Connection conditions 22 7 2 Connecting the measuring device 28 7 3 Hardware settings...

Страница 3: ...rical symbols Symbol Meaning A0011197 Direct current A terminal to which DC voltage is applied or through which direct current flows A0011198 Alternating current A terminal to which alternating voltag...

Страница 4: ...83 Preferred Indicates procedures processes or actions that are preferred A0011184 Forbidden Indicates procedures processes or actions that are forbidden A0011193 Tip Indicates additional information...

Страница 5: ...instructions in the Operating Instructions and supplementary documentation as well as in the certificates depending on the application Following instructions and basic conditions 2 2 Designated use A...

Страница 6: ...ss Hauser is glad to provide assistance in verifying the corrosion resistance of fluid wetted materials but does not accept any warranty or liability as minute changes in the temperature concentration...

Страница 7: ...ndition in which it is safe to operate It meets general safety standards and legal requirements It also complies with the EC directives listed in the device specific EC Declaration of Conformity Endre...

Страница 8: ...4 5 6 7 8 9 E A0014056 1 Important components of a measuring device 1 Electronics compartment cover 2 Display module 3 Main electronics module 4 Cable glands 5 Transmitter housing incl HistoROM 6 I O...

Страница 9: ...0015502 A0013699 Do the nameplate data match the ordering information on the delivery note A0015502 A0013697 Is the CD ROM with the Technical Documentation depends on device version and documents pres...

Страница 10: ...er no 1 2 3 4 A0021952 2 Example of a nameplate 1 Order code 2 Serial number Ser no 3 Extended order code Ext ord cd 4 2 D matrix code QR code For detailed information on the breakdown of the specific...

Страница 11: ...ble both for android smart phones Google Play store and for iPhones and iPads App Store Via the serial number 1 Launch the Endress Hauser Operations App 2 Enter the serial number Ser no of the device...

Страница 12: ...g or slipping Observe the weight specified on the packaging stick on label Observe the transport instructions on the stick on label on the electronics compartment cover A0015606 Observe the following...

Страница 13: ...Proline Promass E 200 Storage and transport Endress Hauser 13 A0015605...

Страница 14: ...by the construction of the device 6 1 1 Mounting position Mounting location A0015595 Installation in down pipes However the following installation suggestion allows for installation in an open vertic...

Страница 15: ...4 5 A0015596 3 Installation in a down pipe e g for batching applications 1 Supply tank 2 Sensor 3 Orifice plate pipe restriction 4 Valve 5 Batching tank DN orifice plate pipe restriction mm in mm in...

Страница 16: ...tion C Horizontal orientation transmitter head down A0015590 2 Exception D Horizontal orientation transmitter head at side A0015592 1 Applications with low process temperatures may reduce the ambient...

Страница 17: ...range Measuring device 40 to 60 C 40 to 140 F Local display 20 to 60 C 4 to 140 F The readability of the display may be impaired at temperatures outside the temperature range If operating outdoors Av...

Страница 18: ...ns Electrical heating e g with electric band heaters Via pipes carrying hot water or steam Via heating jackets For detailed information about heating with electrical band heaters refer to the Operatin...

Страница 19: ...point adjustment is advisable only in special cases To achieve maximum measuring accuracy even with low flow rates Under extreme process or operating conditions e g very high process temperatures or...

Страница 20: ...e gaskets are clean and undamaged Install the gaskets correctly 1 Ensure that the direction of the arrow on the nameplate of the sensor matches the flow direction of the fluid 2 Install the measuring...

Страница 21: ...on document on the CD ROM provided Ambient temperature 17 Measuring range Has the correct orientation for the sensor been selected According to sensor type According to medium temperature According to...

Страница 22: ...s provided by the customer must fulfill the following requirements Electrical safety In accordance with applicable federal national regulations Permitted temperature range 40 C 40 F to 80 C 176 F Mini...

Страница 23: ...20 1 5 with cable 6 to 12 mm 0 24 to 0 47 in Plug in spring terminals for device version without integrated overvoltage protection wire cross sections 0 5 to 2 5 mm2 20 to 14 AWG Screw terminals for d...

Страница 24: ...ission Ground terminal for cable shield Order code for Output Terminal numbers Output 1 Output 2 1 2 3 4 Option A 4 20 mA HART passive Option B 1 4 20 mA HART passive Pulse frequency switch output pas...

Страница 25: ...n the feed side Experience shows that the best results with regard to EMC are achieved in most cases in installations with one sided shielding on the feed side without capacitance termination at the f...

Страница 26: ...tential equalization the multiple grounding of the cable shield causes mains frequency equalizing currents Damage to the bus cable shield Only ground the bus cable shield to either the local ground or...

Страница 27: ...V DC 32 V 1 External supply voltage of the power supply unit with load 2 For device versions with SD03 local display The terminal voltage must be increased by DC 2 V if backlighting is used 3 For dev...

Страница 28: ...1 Remove dummy plug if present 2 NOTICE Insufficient sealing of the housing Operational reliability of the measuring device could be compromised Use suitable cable glands corresponding to the degree...

Страница 29: ...mm in 20mm 3 mm A0013836 Connect the cable in accordance with the terminal assignment For HART communication When connecting the cable shielding to the ground terminal observe the grounding concept o...

Страница 30: ...the device is not recognized by the master All measuring devices are delivered from the factory with the device address 126 and with the software addressing method 2 9 8 3 4 5 6 7 8 1 D IP O N A00156...

Страница 31: ...ch 8 is set to the ON position the address is defined in the operating menu Setup menu Communication submenu Device address parameter 7 4 Ensuring the degree of protection The measuring device fulfill...

Страница 32: ...s installed firmly tightened and leak tight Cable run with water trap 31 Depending on the device version are all the device plugs firmly tightened 28 Does the supply voltage match the specifications o...

Страница 33: ...enu for operators and maintenances Operator Maintenance task oriented function oriente d Expert A0014058 EN 8 Schematic structure of the operating menu 8 1 2 Operating philosophy The individual parts...

Страница 34: ...channel number symbol for diagnostic behavior 1 4 Status area 1 5 Measured value 1 6 Unit for the measured value 1 7 Operating elements 2 Operational display with measured value shown as 1 bar graph...

Страница 35: ...rence density A0013947 Temperature A0013943 Totalizer The measurement channel number indicates which of the three totalizers is displayed A0013945 Output The measurement channel number indicates which...

Страница 36: ...A0013967 A0013968 A0013972 A0013963 Submenu Wizard Parameters within a wizard Parameter locked 8 2 3 Editing view Input mask Operating symbols in the numeric editor A0013985 A0016621 A0013986 Confirms...

Страница 37: ...s the selection bar downwards in a choose list With a Wizard Confirms the parameter value and goes to the next parameter With a text and numeric editor Moves the selection bar to the right forwards in...

Страница 38: ...the keys simultaneously Reduces the contrast brighter setting A0013954 Plus Enter key combination press and hold down the keys simultaneously Increases the contrast darker setting A0013955 Minus Plus...

Страница 39: ...is indicated by the Access status display parameter Navigation path Operation Access status display 8 2 7 Disabling write protection via access code If the symbol appears on the local display in fron...

Страница 40: ...Switching on the keypad lock The keypad lock is switched on automatically Each time the device is restarted If the device has not been operated for longer than one minute in the measured value display...

Страница 41: ...rol system e g PLC 2 Transmitter power supply unit e g RN221N with communication resistor 3 Connection for Commubox FXA195 and Field Communicator 475 4 Field Communicator 475 5 Computer with operating...

Страница 42: ...3 Computer with FieldCare operating tool with COM DTM CDI Communication FXA291 9 System integration For information on system integration see the Operating Instructions for the device 10 10 Commission...

Страница 43: ...with the device 10 3 Setting the operating language Factory setting English or ordered local language XXXX XXX X X 20 50 Operation Setup Main menu 0104 1 Language English Espa ol Fran ais Language En...

Страница 44: ...etup menu Device tag Parameter overview with brief description Parameter Description User entry Factory setting Device tag Enter tag for measuring point Max 32 characters such as letters numbers or sp...

Страница 45: ...operational display mode from the navigation and editing view If write access is activated via access code it can be also be deactivated only via the access code 39 The user role with which the user...

Страница 46: ...Hardware locked option is displayed in the Locking status parameter In addition on the local display the symbol appears in front of the parameters in the header of the operational display and in the n...

Страница 47: ...age A0013940 EN 11 Message for remedial measures 1 Diagnostic information 2 Short text 3 Service ID 4 Diagnostic behavior with diagnostic code 5 Operation time of occurrence 6 Remedial measures The us...

Страница 48: ...www addresses endress com...