9 Trouble-shooting
Proline Promass 83 FOUNDATION Fieldbus
76
Hauser
261
Device status message (FF):
Communication failure –
Err. No. 261
Local display:
S
COMMUNICATION I/O
$
# 261
[.....] Quality = BAD
OUT. QUALITY = BAD
Cause:
Communication error. No data recep-
tion between amplifier and I/O board
or faulty internal data transfer.
Remedy:
Check whether the electronics boards
are correctly inserted in their holders
[.....] Substatus = Device failure
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Device needs
maintenance now
BLOCK_ERR = Input Failure
(faulty input value from Transducer
Block)
Transducer_Error = I/O failure
(input/output error)
379
380
Device status message (FF):
Measuring tube is outside the
range – Err. No. 379 / 380
Local display:
S
FREQ. LIM
$
# 379 / 380
[.....] Quality = BAD
OUT. QUALITY = BAD
Cause:
The measuring tube oscillation fre-
quency is outside the permitted range.
Causes:
– Damaged measuring tube
– Sensor defective or damaged
Remedy:
Contact your Hauser service
organisation.
[.....] Substatus = Sensor failure
OUT. SUBSTATUS = Sensor
Failure
BLOCK_ERR = Device needs
maintenance now
BLOCK_ERR = Input Failure
(faulty input value from Transducer
Block)
Transducer_Error = Mechanical
failure
381
Device status message (FF):
Tube temperatur sensor defect
– Err. No. 381
Local display:
S
FLUIDTEMP.MIN.
$
# 381
[.....] Quality = BAD
OUT. QUALITY = BAD
Cause:
The temperature sensor on the measu-
ring tube is likely defective
Remedy:
Check the following electrical
connections before you contact your
Hauser service organisation:
– Verify that the sensor signal cable
connector is correctly plugged into
the amplifier board
– Remote version:
Check sensor and transmitter termi-
nal connections No. 9 and No. 10
[.....] Substatus = Sensor failure
OUT. SUBSTATUS = Sensor
Failure
BLOCK_ERR = Device needs
maintenance now
BLOCK_ERR = Input Failure
(faulty input value from Transducer
Block)
Transducer_Error = Mechanical
failure
382
Device status message (FF):
Tube temperatur sensor defect
– Err. No. 382
Local display:
S
FLUIDTEMP.MAX.
$
# 382
[.....] Quality = BAD
OUT. QUALITY = BAD
[.....] Substatus = Sensor failure
OUT. SUBSTATUS = Sensor
Failure
BLOCK_ERR = Device needs
maintenance now
BLOCK_ERR = Input Failure
(faulty input value from Transducer
Block)
Transducer_Error = Mechanical
failure
No.
Device status message FF
(Diag. - Act. Syst. Cond.)
Message on local display
Transducer block
Error messages
[.....] = process variable (e.g. mass
flow, volume flow, ...)
Analog Input Funct. Block
Error messages
Cause/remedy
Содержание Proline Promass 83
Страница 24: ...3 Installation Proline Promass 83 FOUNDATION Fieldbus 24 Endress Hauser ...
Страница 34: ...4 Wiring Proline Promass 83 FOUNDATION Fieldbus 34 Endress Hauser ...
Страница 64: ...6 Commissioning Proline Promass 83 FOUNDATION Fieldbus 64 Endress Hauser ...
Страница 66: ...7 Maintenance Proline Promass 83 FOUNDATION Fieldbus 66 Endress Hauser ...
Страница 90: ...9 Trouble shooting Proline Promass 83 FOUNDATION Fieldbus 90 Endress Hauser ...
Страница 116: ...Index Proline Promass 83 FOUNDATION Fieldbus 116 Endress Hauser ...
Страница 118: ...www endress com worldwide BA065D 06 en 12 04 50098623 FM SGML 6 0 ...