9 Trouble-shooting
Hauser
79
588
Device status message (FF):
Overdriving of the analog/digi-
tal converter –
Err. No. 588
Local display:
P
GAIN RED.IMPOS
$
# 588
[.....] Quality = BAD
OUT. QUALITY = BAD
Cause:
Overdriving of the internal analog to
digital converter.
A continuation of the measurement is
no longer possible!
Causes:
– Cavitation
– Extreme pressure pulses
– High gas flow velocity
Remedy:
Change or improve process conditions,
e.g. by reducing the flow velocity.
[.....] Substatus = Sensor failure
OUT. SUBSTATUS = Sensor
Failure
BLOCK_ERR = Device needs
maintenance now
BLOCK_ERR = Input Failure
(faulty input value from Transducer
Block)
Transducer_Error = Mechanical
failure
601
Device status message (FF):
Positive Zero Return active –
Err. No. 601
Local display:
S
POS. ZERO RETURN
!
# 601
[.....] Quality = UNCERTAIN
OUT. QUALITY = UNCERTAIN
Cause:
Positive zero return is active
Remedy:
Deactivate positive zero return:
• FF: Transducer Block
→
Parameter
“System - Positive Zero Return”
→
OFF
• Local display:
BASIC FUNCTIONS
→
SYSTEM
PARAMETERS
→
CONFIGURATI-
ONS
→
POS. ZERO RETURN
(
→
OFF)
[.....] Substatus = Non specific
OUT. SUBSTATUS =
Non specific
–
–
–
691
Device status message (FF):
Simulation of response to error
– Err. No. 691
Local display:
S
SIM. FAILSAFE
!
# 691
[.....] Quality = UNCERTAIN
OUT. QUALITY = UNCERTAIN
Cause:
Simulation of response to error (out-
puts) is active
Remedy:
Switch off simulation:
• FF: not possible
• Local display:
SUPERVISION
→
SYSTEM
→
OPOERATION
→
SIM. FAILSAFE-
MODE (
→
OFF)
[.....] Substatus = Non specific
OUT. SUBSTATUS =
Non specific
BLOCK_ERR = Simulation active
BLOCK_ERR = Simulation active
–
692
Device status message (FF):
Simulation of measuring active
– Err. No. 692
Local display:
S
SIM. MEASURAND
!
# 692
[.....] Quality = UNCERTAIN
OUT. QUALITY = UNCERTAIN
Cause:
Simulation is active
Remedy:
Switch off simulation:
• FF: Transducer Block
→
Parameter
“Simulation -Measurand”
→
OFF
• Vor-Ort-Anzeige:
SUPERVISION
→
SYSTEM
→
OPERATION
→
SIM. MEASU-
RAND (
→
OFF)
[.....] Substatus = Non specific
OUT. SUBSTATUS =
Non specific
BLOCK_ERR = Simulation active
BLOCK_ERR = Simulation active
–
No.
Device status message FF
(Diag. - Act. Syst. Cond.)
Message on local display
Transducer block
Error messages
[.....] = process variable (e.g. mass
flow, volume flow, ...)
Analog Input Funct. Block
Error messages
Cause/remedy
Содержание Proline Promass 83
Страница 24: ...3 Installation Proline Promass 83 FOUNDATION Fieldbus 24 Endress Hauser ...
Страница 34: ...4 Wiring Proline Promass 83 FOUNDATION Fieldbus 34 Endress Hauser ...
Страница 64: ...6 Commissioning Proline Promass 83 FOUNDATION Fieldbus 64 Endress Hauser ...
Страница 66: ...7 Maintenance Proline Promass 83 FOUNDATION Fieldbus 66 Endress Hauser ...
Страница 90: ...9 Trouble shooting Proline Promass 83 FOUNDATION Fieldbus 90 Endress Hauser ...
Страница 116: ...Index Proline Promass 83 FOUNDATION Fieldbus 116 Endress Hauser ...
Страница 118: ...www endress com worldwide BA065D 06 en 12 04 50098623 FM SGML 6 0 ...