Proline Promag P 200
Hauser
5
Measuring system
The device consists of a transmitter and a sensor.
One device version is available: compact version, transmitter and sensor form a mechanical unit.
Transmitter
Promag 200
Materials:
Aluminum coating AlSi10Mg
Configuration:
• External operation via four-line, illuminated local display with touch control
and guided menus ("Make-it-run" wizards) for applications
• Via operating tools (e.g. FieldCare)
A0013471
Sensor
Promag P
Nominal diameter range: DN 15 to 200 (½ to 8")
Materials:
• Sensor housing: aluminum coated AlSi10Mg
• Measuring tubes: stainless steel 1.4301/304, 1.4306/304L
• Liner: PFA, PTFE
• Process connections: 1.0425/316L, 1.4571/316L, A105, C22, F316L, FE
410W B, HII, S235JRG2
• Electrodes: 1.4435/316L, Alloy C-22, tantalum, platinum
• Seals: as per DIN EN 1514-1
• Ground disks: 1.4435/316L, Alloy C-22, tantalum
A0017703
Input
Measured variable
Direct measured variables
Volume flow (proportional to induced voltage)
Calculated measured variables
Mass flow
Measuring range
Typically v = 0.01 to 10 m/s (0.03 to 33 ft/s) with the specified accuracy
Flow characteristic values in SI units
Nominal
diameter
Recommended
flow
Factory settings
min./max. full scale value
(v
~
0.310 m/s/
Full scale value current
output
(v
~
2.5 m/s)
Pulse value
(
~
2 pulse/s)
Low flow cut off
(v
~
0.04 m/s)
[mm]
[in]
[dm
3
/min]
[dm
3
/min]
[dm
3
]
[dm
3
/min]
15
½
4 to 100
25
0.2
0.5
25
1
9 to 300
75
0.5
1
32
–
15 to 500
125
1
2
40
1 ½
25 to 700
200
1.5
3
50
2
35 to 1
100
300
2.5
5
65
–
60 to 2
000
500
5
8
80
3
90 to 3
000
750
5
12
100
4
145 to 4
700
1
200
10
20
125
–
220 to 7
500
1
850
15
30
Содержание Proline Promag P 200
Страница 39: ......