Special mounting instructions
Weather protection cover
To ensure that the optional weather protection cover can be easily opened, maintain the following minimum
head clearance: 350 mm (13.8 in)
Environment
Ambient temperature range
Transmitter
–40 to +60
°
C (–40 to +140
°
F)
Local display
–20 to +60
°
C (–4 to +140
°
F), the readability of the display may be impaired at
temperatures outside the temperature range.
Sensor
• Flange material carbon steel: –10 to +60
°
C (+14 to +140
°
F)
• Flange material stainless steel: –40 to +60
°
C (–40 to +140
°
F)
Liner
Do not exceed or fall below the permitted temperature range of the liner
(
If operating outdoors:
• Install the measuring device in a shady location.
• Avoid direct sunlight, particularly in warm climatic regions.
• Avoid direct exposure to weather conditions.
Weather protection covers can be ordered from Hauser: see "Accessories" section (
Storage temperature
The storage temperature corresponds to the operating temperature range of the measuring transmitter and the
appropriate measuring sensors.
• Protect the measuring device against direct sunlight during storage in order to avoid unacceptably high surface
temperatures.
• Select a storage location where moisture cannot collect in the measuring device as fungus or bacteria
infestation can damage the liner.
• If protection caps or protective covers are mounted these should never be removed before installing the
measuring device.
Degree of protection
Transmitter
• As standard: IP66/67, type 4X enclosure
• When housing is open: IP20, type 1 enclosure
• Display module: IP22, type 1 enclosure