Potential equalization via ground terminal and pipe flanges
Note the following when installing:
• Connect both sensor flanges to the pipe flange via a ground cable and ground them.
• Connect the connection housing of the transmitter or sensor to ground potential by means of the ground
terminal provided for the purpose.To mount the ground cable:
– If DN
£
300 (12"): Mount the ground cable directly on the conductive flange coating of the sensor with
the flange screws.
– If DN
³
350 (14"): Mount the ground cable directly on the metal transport bracket.
Ground cable
Copper wire, at least 6 mm
2
(0.0093 in
2
)
The necessary ground cable can be ordered from Hauser (
Plastic pipe or pipe with insulating liner
This connection method also applies in situations where:
• The customary potential equalization is not used
• Equalizing currents are present
A0016318
å
8
Potential equalization via ground terminal and ground disks
Note the following when installing:
The ground disks must be connected to the ground terminal via the ground cable and be connected to ground
potential.
Ground cable
Copper wire, at least 6 mm
2
(0.0093 in
2
)
The ground cable and ground disks can be ordered from Hauser (