Application

• Accurate bidirectional measurement of liquids with a

minimum conductivity of 

³

 20 

m

S/cm for chemical

applications.

• The electromagnetic measuring principle is unaffected

by pressure and temperature. Additionally the flow
profile has a minimal effect on the measurement
results.

Device properties

• Medium temperature: max. +150 °C (+302 °F)
• Nominal diameter: DN 15 to 200 (1/2 to 8")
• Application-specific liner (PFA, PTFE)
• Two-wire aluminium transmitter
• Graphical local display with operation from the outside

(Touch Control)

• Communication via 4-20 mA HART
• Ex approvals accepted worldwide: ATEX, IECEx

cCSAus (intrinsic safety or flameproof enclosure)

Your benefits

Genuine loop-powered technology for seamless 2-wire
integration and robustness in standard process
applications

Sizing 

-

 correct product selection

Applicator

 - the reliable, easy-to-use tool for selecting and

sizing measuring devices for every application

Installation 

-

 simple and efficient

• Compact design
• Suitable for installations in the hazardous area
• Reduced wiring effort due to two-wire technology

Commissioning 

-

 reliable and intuitive

Guided parameterization 

-

 "Make-it-run" wizards

Operation 

-

 increased measurement availability

• Measurement of volume flow
• No pressure loss, no moving parts, immune to

vibrations

• Diagnostics; Automatic data restore by HistoROM
Cost-effective Life Cycle Management by W@M

Technical Information

Proline Promag P 200

Electromagnetic flow measuring system
The device with genuine two-wire technology and for minimized
cost of ownership

TI01062D/06/EN/01.12
71185911

Содержание Proline Promag P 200

Страница 1: ...d technology for seamless 2 wire integration and robustness in standard process applications Sizing correct product selection Applicator the reliable easy to use tool for selecting and sizing measurin...

Страница 2: ...0 Special mounting instructions 21 Environment 21 Ambient temperature range 21 Storage temperature 21 Degree of protection 21 Shock resistance 22 Vibration resistance 22 Mechanical load 22 Electromagn...

Страница 3: ...other connections A0011201 Equipotential connection A connection that has to be connected to the plant grounding system This may be a potential equalization line or a star grounding system depending...

Страница 4: ...he electromagnetic measuring principle the flowing medium is the moving conductor The voltage induced is proportional to the flow velocity and is supplied to the amplifier by means of two measuring el...

Страница 5: ...A105 C22 F316L FE 410W B HII S235JRG2 Electrodes 1 4435 316L Alloy C 22 tantalum platinum Seals as per DIN EN 1514 1 Ground disks 1 4435 316L Alloy C 22 tantalum A0017703 Input Measured variable Dire...

Страница 6: ...output v 2 5 m s Pulse value 2 pulse s Low flow cut off v 0 04 m s in mm gal min gal min gal gal min 15 1 0 to 27 6 0 05 0 10 1 25 2 5 to 80 18 0 2 0 25 32 4 to 130 30 0 2 0 5 1 40 7 to 190 50 0 5 0 7...

Страница 7: ...ble 0 to 999 s Pulse pause ratio 1 1 Assignable measured variables Volume flow Mass flow Switch output Switching behavior Binary conductive or non conductive Switching delay Adjustable 0 to 100 s Numb...

Страница 8: ...MUR recommendation NE 107 Operating tool Via digital communication HART protocol Via service interface Plain text display With information on cause and remedial measures Additional information on remo...

Страница 9: ...19 V Maximum load RB 19 V 13 5 V 0 022 A 250 W Ex connection data Safety related values Ex d type of protection Order code for Output Output type Safety related values Option A 4 20 mA HART Unom DC 3...

Страница 10: ...code for Output Output type Intrinsically safe values Option A 4 20 mA HART Ui DC 30 V Ii 300 mA Pi 1 W Li 0 mH Ci 5 nF Option B 4 20 mA HART Ui DC 30 V Ii 300 mA Pi 1 W Li 0 mH Ci 5 nF Pulse frequen...

Страница 11: ...otection Maximum number of terminals with integrated overvoltage protection 1 2 3 Output 1 passive supply voltage and signal transmission Output 2 passive supply voltage and signal transmission Ground...

Страница 12: ...n A 4 20 mA HART 770 mW Option B 4 20 mA HART pulse frequency switch output Operation with output 1 770 mW Operation with output 1 and 2 2770 mW For information on the Ex connection values 9 Current c...

Страница 13: ...um load 8 5 Connection for HART operating devices 33 6 Analog display unit observe maximum load 8 7 Transmitter 2 4 20 mA 3 1 4 A0015512 3 Connection example for 4 20 mA current output passive 1 Autom...

Страница 14: ...ensure correct measurement Same electrical potential for the fluid and sensor Company internal grounding concepts Pipe material and grounding Connection examples for standard situations Metal grounded...

Страница 15: ...If DN 350 14 Mount the ground cable directly on the metal transport bracket Ground cable Copper wire at least 6 mm2 0 0093 in2 The necessary ground cable can be ordered from Endress Hauser 36 Plastic...

Страница 16: ...Guide the shield of the signal lines through a capacitor The necessary ground cable can be ordered from Endress Hauser 36 Terminals For device version without integrated overvoltage protection plug i...

Страница 17: ...class restrictions apply to the ambient temperature for device versions with overvoltage protection Performance characteristics Reference operating conditions To DIN EN 29104 Fluid temperature 28 2 C...

Страница 18: ...ture coefficient at zero point 4 mA 0 02 10 K max 0 35 over the entire temperature range 40 to 60 C 40 to 140 F Temperature coefficient with span 20 mA 0 05 10 K max 0 5 over the entire temperature ra...

Страница 19: ...lly filled pipes A partially filled pipe with a gradient necessitates a drain type configuration The empty pipe detection EPD function offers additional protection by detecting empty or partially fill...

Страница 20: ...ion Inlet and outlet runs If possible install the sensor upstream from fittings such as valves T pieces or elbows Observe the following inlet and outlet runs to comply with accuracy specifications Inl...

Страница 21: ...ng outdoors Install the measuring device in a shady location Avoid direct sunlight particularly in warm climatic regions Avoid direct exposure to weather conditions Weather protection covers can be or...

Страница 22: ...rovided in the Declaration of Conformity Process Medium temperature range 20 to 150 C 4 to 302 F for PFA 40 to 130 C 40 to 266 F for PTFE n PFA PTFE 0 0 0 0 20 40 40 20 20 40 100 60 140 F C 40 60 80 1...

Страница 23: ...0 60 80 100 120 140 160 180 C 360 F 0 40 100 200 300 200 100 400 300 500 600 0 A0005594 13 Materials C22 FE 410W B and S235JRG2 Flange connection according to EN 1092 1 DIN 2501 PN25 PN16 PN10 PN40 PN...

Страница 24: ...26 15 Material A105 Flange connection according to ASME B16 5 Class300 Class150 0 10 20 30 40 50 bar psi 40 20 0 20 40 60 80 100 120 140 160 180 C 360 F 0 40 100 200 300 200 100 400 300 500 600 700 80...

Страница 25: ...er of the pipe and the flow rate determine the nominal diameter of the sensor The optimum velocity of flow is between 2 to 3 m s 6 56 to 9 84 ft s Also match the velocity of flow v to the physical pro...

Страница 26: ...94 Vibrations In the event of very strong vibrations the pipe and sensor must be supported and fixed For information on the permitted resistance to vibration and shock 22 22 L A0016266 18 Measures to...

Страница 27: ...60 165 75 90 180 189 5 115 94 305 109 414 80 200 162 102 60 165 75 90 180 189 5 115 94 305 109 414 100 250 162 102 60 165 75 90 180 189 5 115 94 305 109 414 125 250 162 102 60 165 75 90 260 189 5 155...

Страница 28: ...6 5 91 19 5 6 11 8 6 38 4 02 2 36 6 50 2 95 3 54 10 2 7 46 6 10 5 51 13 6 5 91 19 5 8 13 8 6 38 4 02 2 36 6 50 2 95 3 54 12 8 7 46 7 09 6 14 14 6 7 09 21 7 1 The length L is always the same and does n...

Страница 29: ...gs and without packaging material Weight in SI units Nominal diameter EN DIN AS 1 ASME JIS mm in Pressure rating kg Pressure rating kg Pressure rating kg 15 PN 40 5 0 Class 150 5 0 10K 5 0 25 1 PN 40...

Страница 30: ...PN 16 Class 150 10K 75 2 95 80 3 15 100 4 PN 16 Class 150 10K 101 3 98 104 4 09 125 PN 16 10K 126 4 96 129 5 08 150 6 PN 16 Class 150 10K 154 6 06 156 6 14 200 8 PN 10 Class 150 10K 201 7 91 202 7 95...

Страница 31: ...s reference electrodes and electrodes for empty pipe detection Standard 1 4435 316L Alloy C 22 platinum tantalum Optional only measuring electrodes made from platinum Process connections EN 1092 1 DIN...

Страница 32: ...ode for Display Operation option C SD02 Order code for Display Operation option E SD03 E 1 A0015544 1 E A0015546 1 Operation with pushbuttons 1 Operation with touch control Display elements 4 line dis...

Страница 33: ...ubox FXA195 and Field Communicator 475 4 Field Communicator 475 5 Computer with operating tool e g FieldCare AMS Device Manager SIMATIC PDM 6 Commubox FXA195 USB 7 Field Xpert SFX100 8 VIATOR Bluetoot...

Страница 34: ...cument Reference is made to this document on the nameplate The separate Ex documentation XA containing all the relevant explosion protection data is available from your Endress Hauser sales center ATE...

Страница 35: ...NE 107 Self monitoring and diagnosis of field devices NAMUR NE 131 Requirements for field devices for standard applications Ordering information Detailed ordering information is available from the fo...

Страница 36: ...r code to define the following specifications Approvals Output Display operation Housing Software For details see Installation Instructions EA00104D Weather protection cover Is used to protect the mea...

Страница 37: ...Software for selecting and sizing Endress Hauser measuring devices Calculation of all the necessary data for identifying the optimum flowmeter e g nominal diameter pressure loss accuracy or process co...

Страница 38: ...acks For details see Technical Information TI00081R and Brief Operating Instructions KA00110R Documentation The following document types are available On the CD ROM supplied with the device In the Dow...

Страница 39: ......

Страница 40: ...national Endress Hauser Instruments International AG Kaegenstrasse 2 4153 Reinach Switzerland Tel 41 61 715 81 00 Fax 41 61 715 25 00 www endress com info ii endress com TI01062D 06 EN 01 12 71185911...

Отзывы: