Proline Promag 55 FOUNDATION Fieldbus
Commissioning
Hauser
69
!
Note!
• The display returns to the SETUP COMMISSIONING (1002) cell if you press the
X
key
combination during parameter interrogation. The stored parameters remain valid.
• The system units selected via the Quick Setup are only valid for the local display and for
parameters in the Transducer Blocks. They do not affect the process variables which are
transmitted via FOUNDATION Fieldbus.
m
The DELIVERY SETTINGS option sets each selected unit to the factory setting.
The ACT.SETTING option accepts the units previously set by you.
n
Only units not yet configured in the current setup are offered for selection in each cycle. The unit for Mass and
Volume is derived from the corresponding flow unit.
o
The "YES" option remains visible until all the units have been configured.
"NO" is the only option displayed when no further units are available.
p
The "automatic parameterization of the display" option contains the following basic settings/factory settings:
YES
Main line = Volume flow
Additional line = Totalizer 1
Information line = Operating/System conditions
NO
The existing (selected) settings remain.