Proline Promag 55 FOUNDATION Fieldbus
Installation
Hauser
27
Installing the high-temperature version (with PFA lining)
The high-temperature version has a housing support for the thermal separation of sensor and
transmitter. The high-temperature version is always used for applications in which high ambient
temperatures are encountered in conjunction with high fluid temperatures. The high-temperature
version is obligatory if the fluid temperature e150 °C (+300 °F).
!
Note!
You will find information on permissible temperature ranges on
Insulation
Pipes generally have to be insulated if they carry very hot fluids to avoid energy losses and prevent
accidental contact with pipes at temperatures that could cause injury. Guidelines regulating the
insulation of pipes have to be taken into account.
"
Caution!
Risk of electronics overheating. The housing support dissipates heat and its entire surface area must
remain uncovered. Make sure that the sensor insulation does not extend past the top of the two
sensor half-shells.
a0004300
Fig. 19:
Sensor (high-temperature version): insuIating the pipe
Esc
E
-
+
max.
Содержание Proline Promag 55
Страница 2: ......
Страница 120: ...www endress com worldwide BA126D 06 EN 07 09 71089883 FM SGML6 0 ProMoDo ...