Proline Promag 50H, 53H
20
Hauser
Pressure loss
• With DN 8 to 100 (3/8 to 4") no pressure loss if the sensor is installed in a pipe with the same nominal
diameter.
• Pressure losses for configurations incorporating adapters according to DIN EN 545
16, Section "Adapters").
Flow characteristic values (US units)
Diameter
Recommended flow rate
Factory settings
[inch]
[mm]
Min./max. full scale value
(v ~ 0.3 or 10 m/s)
Full scale value, current output
(v ~ 2.5 m/s)
Pulse value
(~ 2 pulses/s)
Low flow cut off
(v ~ 0.04 m/s)
1/12"
2
0.015 to 0.5 gal/min
0.1 gal/min
0.001 gal
0.002 gal/min
1/8"
4
0.07 to 2 gal/min
0.5 gal/min
0.005 gal
0.008 gal/min
3/8"
8
0.25 to 8 gal/min
2 gal/min
0.02 gal
0.025 gal/min
½"
15
1.0 to 27 gal/min
6 gal/min
0.05 gal
0.10 gal/min
1"
25
2.5 to 80 gal/min
18 gal/min
0.20 gal
0.25 gal/min
1½"
40
7 to 190 gal/min
50 gal/min
0.50 gal
0.75 gal/min
2"
50
10 to 300 gal/min
75 gal/min
0.50 gal
1.25 gal/min
3"
80
24 to 800 gal/min
200 gal/min
2.00 gal
2.50 gal/min
4"
100
40 to 1250 gal/min
300 gal/min
2.00 gal
4.00 gal/min
Содержание Proline Promag 50H