Proline Promag 300 Modbus RS485
Description of device parameters
Hauser
207
Puls.outp.sim. 1 to n
Navigation
Expert → Diagnostics → Simulation → Puls.outp.sim. 1 to n (0458–1 to n)
Prerequisite
In the
Operating mode
Pulse
option is selected.
Description
Use this function to switch simulation of the pulse output on and off. The display
alternates between the measured value and a diagnostic message of the "Function check"
category (C) while simulation is in progress.
Selection
• Off
• Fixed value
• Down-count. val.
Factory setting
Off
Additional information
Description
The desired simulation value is defined in the
Pulse value 1 to n
parameter.
Selection
• Off
Pulse simulation is switched off. The device is in normal measuring mode or another
process variable is being simulated.
• Fixed value
Pulses are continuously output with the pulse width specified in the
Pulse width
• Down-count. val.
The pulses specified in the
Pulse value
parameter (→ 207) are output.
Pulse value 1 to n
Navigation
Expert → Diagnostics → Simulation → Pulse value 1 to n (0459–1 to n)
Prerequisite
In the
Puls.outp.sim. 1 to n
parameter, the
Down-count. val.
option is selected.
Description
Use this function to enter a pulse value for the simulation. In this way, users can verify the
correct adjustment of the pulse output and the correct function of downstream switching
units.
User entry
0 to 65
535
Switch sim. 1 to n
Navigation
Expert → Diagnostics → Simulation → Switch sim. 1 to n (0462–1 to n)
Prerequisite
In the
Operating mode
Switch
option is selected.