Proline Promag 300 Modbus RS485
Description of device parameters
Hauser
101
Additional information
Current input behavior
The current input behaves differently depending on the settings configured in the
following parameters:
• Failure mode (→ 101)
Configuration examples
Pay attention to the configuration examples for
4 mA value
20 mA value
Navigation
Expert → Input → Current input 1 to n → 20 mA value (1607–1 to n)
Description
Use this function to enter a value for the 20 mA current.
User entry
Signed floating-point number
Factory setting
Depends on country and nominal diameter
Additional information
Configuration examples
Pay attention to the configuration examples for
4 mA value
Failure mode
Navigation
Expert → Input → Current input 1 to n → Failure mode (1601–1 to n)
Description
Use this function to select the input behavior when measuring a current outside the
configured
Current span
Selection
• Alarm
• Last valid value
• Defined value
Factory setting
Alarm
Additional information
Options
• Alarm
An error message is set.
• Last valid value
The last valid measured value is used.
• Defined value
A user-defined measured value is used (
Failure value