background image

Products

Solutions

Services

Brief Operating Instructions

Flowmeter

Proline 10

Modbus RS485 transmitter

Transmitter with Coriolis sensor

These instructions are Brief Operating Instructions; they are

not

 a substitute for the Operating Instructions pertaining to

the device.

Brief Operating Instructions part 2 of 2: Transmitter

Contain information about the transmitter.
Brief Operating Instructions part 1 of 2: Sensor →   3

KA01500D/06/EN/01.21

71510510

2021-07-01

Содержание Proline 10

Страница 1: ...s sensor These instructions are Brief Operating Instructions they are not a substitute for the Operating Instructions pertaining to the device Brief Operating Instructions part 2 of 2 Transmitter Contain information about the transmitter Brief Operating Instructions part 1 of 2 Sensor 3 KA01500D 06 EN 01 21 71510510 2021 07 01 ...

Страница 2: ...owmeter Proline 10 Modbus RS485 2 Endress Hauser Order code Ext ord cd Ser no www endress com deviceviewer Endress Hauser Operations App XXXXXXXXXXXX XXXXX XXXXXX XXX XXXX XX Serial number 1 3 2 A0023555 ...

Страница 3: ...easuring device Incoming acceptance and product identification Storage and transport Installation Brief Operating Instructions Part 2 Transmitter The Transmitter Brief Operating Instructions are aimed at specialists with responsibility for commissioning configuring and parameterizing the measuring device until the first measured value Product description Installation Electrical connection Operatio...

Страница 4: ...Product design 9 4 Installation 9 4 1 Turning the transmitter housing 10 4 2 Post installation check 11 5 Electrical connection 12 5 1 Connection conditions 12 5 2 Requirements for connecting cable 12 5 3 Connecting the transmitter 14 5 4 Removing a cable 17 5 5 Ensuring potential equalization 17 5 6 Hardware settings 18 5 7 Post connection check 19 6 Operation 20 6 1 Overview of the operating opt...

Страница 5: ...potentially dangerous situation Failure to avoid the situation may result in a minor or mild injury NOTICE This symbol alerts you to a potentially harmful situation Failure to avoid the situation may result in damage to the facility or to something in the facility s vicinity 1 1 2 Electronics Direct current Alternating current Direct current and alternating current Protective earthing 1 1 3 Device...

Страница 6: ...e of the device must only be carried out by trained specialist personnel authorized by the facility s owner operator Before commencing work the trained specialist personnel must carefully read understand and adhere to the Operating Instructions additional documentation and certificates Comply with national regulations 2 2 Requirements for operating personnel Operating personnel are authorized by t...

Страница 7: ...he flange Observe tightening torques 2 8 Electrical connection Comply with national installation regulations and guidelines Observe cable specifications and device specifications Check the cable for damage If using the device in hazardous areas observe the Safety Instructions documentation Provide establish potential equalization Provide establish grounding 2 9 Surface temperature Media with eleva...

Страница 8: ...asurement of liquids and gases Depending on the version ordered the device measures potentially explosive flammable poisonous and oxidizing media Devices for use in a hazardous area in hygienic applications or where there is an increased risk due to process pressure are labeled accordingly on the nameplate Non designated use can compromise safety The manufacturer is not liable for damage caused by...

Страница 9: ...ign The transmitter and sensor form a mechanical unit 4 1 2 3 A0043525 1 Main device components 1 Housing cover 2 Display module 3 Transmitter housing 4 Sensor 4 Installation For detailed information about installing the sensor see the Brief Operating Instructions for the sensor 3 ...

Страница 10: ...1 Loosen the fixing screws on both sides of the transmitter housing 2 NOTICE Overrotation of the transmitter housing Interior cables are damaged Turn the transmitter housing a maximum of 180 in each direction Turn the transmitter housing to the desired position 3 Tighten the screws in the logically reverse sequence ...

Страница 11: ...he device comply with the measuring point specifications For example Process temperature Process pressure Ambient temperature Measuring range Has the correct orientation been selected for the device Does the direction of the arrow on the device match the flow direction of the medium Is the device protected against precipitation and sunlight ...

Страница 12: ... measures are required Set up a disconnecting device switch or power circuit breaker to easily disconnect the device from the supply voltage In addition to the device fuse include an overcurrent protection unit with max 10 A in the facility installation Plastic sealing plugs act as safeguards during transportation and must be replaced by suitable individually approved installation material Connect...

Страница 13: ...table for the minimum temperatures and maximum temperatures to be expected 5 2 3 Power supply cable incl conductor for the inner ground terminal A standard installation cable is sufficient Provide grounding according to applicable national codes and regulations 5 2 4 Signal cable Modbus RS485 Cable type A according to EIA TIA 485 standard is recommended Current output 4 to 20 mA Standard installat...

Страница 14: ...al assignment is available Modbus RS485 and current output 4 to 20 mA active Supply voltage Output 1 Output 2 1 2 26 27 24 25 22 B 23 A L N Current output 4 to 20 mA active Modbus RS485 Modbus RS485 and current output 4 to 20 mA passive Supply voltage Output 1 Output 2 1 2 26 27 24 25 22 B 23 A L N Current output 4 to 20 mA passive Modbus RS485 5 3 3 Wiring the transmitter Use a suitable cable gla...

Страница 15: ...ing to the degree of protection A0045442 1 Ground the device carefully and provide potential equalization 2 Connect protective earthing to the outer ground terminals 1 5 Nm 1 1 lbf ft 3 mm A0041094 3 Loosen the Allen key of the securing clamp 4 Unscrew the housing cover in the counterclockwise direction A0041330 5 Press the tab of the display module holder 6 Remove the display module from the disp...

Страница 16: ... Feed the power supply cable and signal cable through the corresponding cable entry 10 0 4 mm in A0041357 10 Strip the cable and cable ends 11 Fit ferrules over the strands and press in place A0041358 The terminal assignment is documented on an adhesive label 12 Connect the protective ground PE to the inner ground terminal 13 Connect the power supply cable and signal cable as per the terminal assi...

Страница 17: ...1 Use a flat blade screwdriver to press down on the slot between the two terminal holes and hold 2 Remove the cable end from the terminal 5 5 Ensuring potential equalization No special measures for potential equalization are required For devices intended for use in hazardous locations please observe the guidelines in the Ex documentation XA ...

Страница 18: ... of the securing clamp 2 Unscrew the housing cover in the counterclockwise direction A0041330 3 Press the tab of the display module holder 4 Remove the display module from the display module holder On 1 A0044412 5 Set the write protection switch on the back of the display module to the On position Write protection is enabled 6 Follow the sequence in the reverse order to reassemble ...

Страница 19: ...et the requirements Is the terminal assignment correct Are all the cable glands installed firmly tightened and leak tight Are dummy plugs inserted in unused cable entries Are transportation plugs replaced by dummy plugs Are the housing screws and housing cover tightened Do the cables loop down before the cable gland water trap A0042316 Does the supply voltage match the specifications on the transm...

Страница 20: ...rating tool e g FieldCare DeviceCare AMS Device Manager SIMATIC PDM 3 Field Xpert SFX350 or SFX370 via Bluetooth 4 Field Xpert SMT70 via Bluetooth 5 Tablet or smartphone via Bluetooth 6 Automation system e g PLC 6 2 Local operation 6 2 1 Unlocking local operation Local operation must first be unlocked before the device can be operated via the touch screen To unlock draw the pattern L on the touch ...

Страница 21: ...routine operation the local display shows the operational display screen The operational display consists of several windows which the user can toggle between The operational display can be customized see the description of parameters 23 Operational display and navigation V 3 23 23 0 3 23 8 3 kg min kg C 1 2 Ʃ2 T3 ṁ m min 3 XXXXX 3 4 A0042992 1 Quick access 2 Status symbols communication symbols a...

Страница 22: ...lection of specific device functions Quick access is indicated by a triangle at the top of the local display in the middle Quick access and navigation 3 23 23 0 3 23 8 3 kg min m min kg C Quick links Bluetooth User manag Actual diag Locking stat 2 1 V Ʃ2 T3 ṁ Lock Screen XXXXX A0044208 1 Quick access 2 Quick access with specific device functions Tap Back to operational display Open specific device...

Страница 23: ...3 23 8 3 kg min m min kg C V Ʃ2 T3 ṁ XXXXX Main menu Diagnostics Application Guidance System User manag Language A A0044213 1 Open the main menu 2 Open menus for the specific device functions Tap Back to operational display Open menus Symbols Back to operational display Guidance menu Configuration of the device Diagnostics menu Troubleshooting and control of device behavior Application menu Applic...

Страница 24: ...enus or parameters Select options Skip items in list Swipe vertically Select items in a list on a step by step basis Symbols Return to previous menu Skip to bottom of list Skip to top of list 6 2 6 Diagnostic information Diagnostic information displays additional instructions or background information for diagnostic events Opening the diagnostic message The diagnostic behavior is indicated on the ...

Страница 25: ...ooting measures 6 2 7 Editing view Editor and navigation The text editor is used to enter characters Text editor Numeric editor A B D C A Z 0 9 a z 6 4 5 3 2 E XXXXX 1 A0043020 1 Entry display area 2 Delete character 3 Confirm your entry 4 Switch input field 5 Cancel editor 6 Input field 1 2 5 4 3 2 3 0 1 4 5 A0043023 1 Entry display area 2 Delete character 3 Confirm your entry 4 Cancel editor 5 I...

Страница 26: ...r Tap Enter characters Select next character set Swipe horizontally Display next or previous page Input field A Upper case a Lower case 1 Numbers Special characters 6 2 8 Date The device has a real time clock for all log functions The time can be configured here ...

Страница 27: ... Cancel editor 6 3 SmartBlue App The device has a Bluetooth interface and can be operated and configured using the SmartBlue App The SmartBlue App must be downloaded onto a terminal device for this purpose Any terminal device can be used The range is 20 m 65 6 ft under reference conditions Incorrect operation by unauthorized persons is prevented by means of encrypted communication and password enc...

Страница 28: ...d information on system integration see the Operating Instructions for the device Overview of device description files Current version data for the device Operating tools Compatibility with previous model Modbus RS485 information Function codes Response time Modbus data map 8 Commissioning 8 1 Post installation check and post connection check Before commissioning the device make sure that the post...

Страница 29: ...ng on the device Switch on the supply voltage for the device The local display switches from the start screen to the operational display Volume flow 1767 2063 l h V Starting up XXXXX A0042938 If device startup is not successful the device displays an error message to this effect ...

Страница 30: ...kg min m min kg C V Ʃ2 T3 ṁ XXXXX Main menu Diagnostics Application Guidance System User manag Language A A0042939 2 Via the Language symbol select the desired language Main menu Diagnostics Application Guidance System User manag Language A A0042940 3 Via the Guidance symbol open the Commissioning wizard Main menu Guidance Commissioning Main menu Diagnostics Application Guidance System User manag ...

Страница 31: ...the SmartBlue App 27 Connecting the SmartBlue App to the device 1 Enable Bluetooth on the mobile handheld terminal tablet or smartphone 2 Start the SmartBlue App A Live List shows all the devices available 3 Select the desired device The SmartBlue App shows the device login 4 Under user name enter admin 5 Under password enter the device s serial number See nameplate for serial number 6 Confirm you...

Страница 32: ... against unauthorized access see the Operating Instructions for the device 9 Diagnostics and troubleshooting 9 1 Diagnostic information on local display 9 1 1 Diagnostic message The local display alternates between displaying faults as a diagnostic message and displaying the operational display screen 3 23 23 0 3 23 8 3 kg min m min kg C V Ʃ2 T3 ṁ XXXXX F180 Temperature exceeded Help 2 4 3 1 XXXXX...

Страница 33: ......

Страница 34: ......

Страница 35: ......

Страница 36: ...www addresses endress com 71510510 71510510 ...

Отзывы: