Diagnostics and troubleshooting
10�4�2 Previous diagnostics
Diagnostics
1000
→
Previous diagnostics
1200
This function is used to display the previous error (code with short message).
10�4�3 Erase previous diagnostics
Diagnostics
1000
→
Erase previous diagnostics
1300
This function is used to erase the previous logged error (code with short message).
Options: "Keep", "Erase"
10.5 Overview of diagnostic information
Code Help text
Remedial action
0000 No error
−
F041 Connection FDR
• Check electrical connection to the FDR57
• Check function of the FDR57, if necessary
• In case of defect: contact service
F042 Connection FQR
• Check electrical connection to the FQR57
• Check function of the FQR57, if necessary
• In case of defect: contact service
F043 Communication FDR
• Check function of the FDR57
• Reset device, if necessary
• In case of defect: contact service
F044 Communication FQR
• Check function of the FQR57
• Reset device, if necessary
• In case of defect: contact service
F261 Flash FTR
• Perform reset of the FTR525
• If the defect occurs repetitively: contact service
F262 EEPROM FTR
F263 EEPROM FQR
• Perform reset of the FQR57
• If the defect occurs repetitively: contact service
F264 EEPROM FDR
• Perform reset of the FDR57
• If the defect occurs repetitively: contact service
F401 FTR: T < Min / T > Max
• Minimum temperature on the device undercut or
maximum temperature exceeded (Technical Data
→
57)
• Ensure operating temperature range (Installation
conditions →
13)
F402 FDR: T < Min / T > Max
F403 FQR: T < Min / T > Max
C482 Simulation
• Only for information, no error
• Deactivate simulation (→
41)
10�6 Reset device
Reset functions of the FTR525
→ 40