Nanomass Gas Density
System integration
Hauser
35
9.2.2
Data structure and commands
The commands which can be used to read measured variables and device information out of
the measuring device are listed below. The data are transmitted in packets. The data are
divided into seven packets. Only packet 5 is variable and can be used. For data interrogation,
the bytes must be specified in hexadecimal format.
Example
String entry to call up measured data: 0x01 0x00 0x00 0x00 0x08 0x02 0x03.
The device returns the following data. Please note that the string values used are only
sample values:
A0027448
Fig. 12: String
entry
1
Separator
2
String name
3
String value
Packet
Byte (hex)
Description
String output
String
name
Description
1
0x01
No access possible
--
2
0x00
No access possible
--
3
0x00
No access possible
--
4
0x00
No access possible
--
5
0x05
Device information
pcb
No function
mtf
No function
den
No function
de
No function
fid
No function
pip
No function
cus
Tag name
(defined by customer)
ser
Serial number
0x08
Measured data
0008
Temperature in °C
0005
Density in kg/m
3
0006
Reference density
0011
Concentration
0009
Pressure in bar absolute
6
0x02
No access possible
--
7
0x03
No access possible
--
@!000820.55 !00050.00062@!00060.00000@!00091.02@!
@
1
2
3
1
2
3
1
2
3
1
2
3
1