Commissioning
Nanomass Gas Density
38
Hauser
10.4.5 Configuring the current output
Various parameters are used to configure the current output. In addition, there are
differences depending on which current output is being configured.
Navigation
1. Menu "Measuring variables"
→
"System units"
2. Menu "Output 4-20 mA"
→
"Output channel 1" and "Output channel 2"
3. Menu "Output 4-20 mA"
→
"Response channel 1/2"
4. Menu "Supervision"
→
"Assign error prompt for channel 2"
Parameter overview with brief description
Menu "Measuring variables"
→
"System units"
Parameter
Procedure
Selection/input
Factory setting
Density
Select the unit for density.
Result
The selected unit applies for:
• The "Density" process variable
• The "4mA Ch1" parameter
• The "20mA Ch1" parameter
• The "4mA Ch2" parameter
• The "20mA Ch2" parameter
Note
If the unit is changed, configured current
output values are automatically adapted
to the new unit.
Specific Gravity = measured density/
density of air;
Molecular weight = average molar mass
according to the ideal gas law
• g/cm³
• g/cc
• kg/L
• kg/m³
• Lb/ft³
• Lb/gal
•
Unit defined under "Text arb.
density"
• Specific gravity
• Molecular weight
kg/m³
Reference density
Select the unit for reference density.
Result
The selected unit applies for: the
"Reference Density" process variable
Note
If the unit is changed, configured current
output values are automatically adapted
to the new unit.
• g/cm³
• g/cc
• kg/L
• kg/m³
• Lb/ft³
• Lb/gal
kg/m³
Temperature
Select the unit for temperature.
Result
The selected unit applies for:
• The "Temperature" process variable
• The "4mA Ch1" parameter
• The "20mA Ch1" parameter
• The "4mA Ch2" parameter
• The "20mA Ch2" parameter
Note
If the unit is changed, configured current
output values are automatically adapted
to the new unit.
• °C
• K
• °F
• °R
Country-specific:
• °C
• °F